background image

 

IDEXX VetLab 

UA

 

Analyzer

 

Operator’s Guide

 13

The LED emits light of a defined wavelength onto the surface of the test pad at an optimum 
angle. The light hitting the test zone is reflected more or less intensely depending on the color 
produced on the test pad, and is picked up by the detector, a phototransistor positioned directly 
above the test zone. The phototransistor sends an analog electrical signal to an A/D converter, 
which changes it to digital form. The microprocessor then converts this digital reading to a relative 
reflectance value by referring it to a calibration standard. 

Finally, the analyzer compares the reflectance value with the defined range limits (reflectance 
values that are programmed into the analyzer for each parameter) and outputs a semi-quantitative 
result. 

Installing the IDEXX VetLab UA Analyzer

1.  Unpack the IDEXX VetLab UA Analyzer and place it on a stable, level 

surface. Choose a well-ventilated area away from obvious sources 
of heat, direct sunlight, cold, humidity, or vibrations. For optimum 
results, room temperature should be at 68°F–78°F (20°C–26°C) and 
relative humidity at 30%–60%. 

IMPORTANT: 

Allow the UA analyzer to acclimate to room 

temperature for two to three hours before operating it for the first time. 

2.  Connect the power cord and the AC adapter. Then, connect the AC 

adapter to the power socket at the rear of the analyzer and plug the 
power cord into an outlet. 

 

NOTE:

 Use 

only

 the power adapter supplied (which is marked as 

supplying 7.5 V DC, 3000 mA). The cords from other power adapters 
may fit in the power socket of the UA analyzer, but may have the 
incorrect voltage or current for this analyzer. Using a different power 
adapter than the one supplied may impact results or render the 
analyzer unusable.

3.  Hold the test strip tray with the gray reference pad facing up (Figure 

1). Slide it into the slot below the function keys so that the edge of the 
tray is flush with the edge of the analyzer (Figure 2). 

 

IMPORTANT:

 Be careful not to touch the gray reference pad. 

Contamination of the reference pad may impair the quality of the 
results.

4.  Load the printer paper. To release the printer cover, press the raised 

dots immediately below the cover (Figure 3). After the cover pops 
open, lift it back. 

 

Place the paper roll in the compartment and pull out the first few 
inches of paper to just beyond the edge of the compartment. The 
thermosensitive side of the paper (the outer surface of the paper roll) should be facing 
downward (Figure 4). 

 

Close the cover by pressing it until it locks audibly.  

5.  Select the printing, units, and connection (interface) options (for more information, see pages 

20–21).

 

NOTE:

 If your UA analyzer is connected to the IDEXX VetLab Station, you can turn off the 

analyzer’s print option, but you must still have paper in the analyzer. UA analyzer results 
will be transferred to the IDEXX VetLab Station automatically and included in the patient’s 
integrated diagnostic report.

Figure 2

Figure 1

Figure 3

Figure 4

gray reference pad

Summary of Contents for VETLAB UA

Page 1: ...IDEXX VetLab UA Analyzer Operator s Guide English Version...

Page 2: ......

Page 3: ...s copyrights or other intellectual or industrial property rights covering this document or subject matter in this document The furnishing of this document does not give a license to these property rig...

Page 4: ......

Page 5: ...s on the Analyzer 19 Overview 19 Using the Memory Options 19 Restarting the Sample Numbering 20 Modifying the Interface Options 20 Printing Settings 20 Selecting Units of Measurement 21 Changing the R...

Page 6: ......

Page 7: ...Ink markers t Sprays disinfectants containing volatile liquids except isopropyl alcohol t Insecticides t Polish t Room freshener t Use only the power adapter supplied which is marked as supplying 7 5...

Page 8: ...rie Consult instructions for use Consulter la notice d utilisation Gebrauchsanweisung beachten Consultare le istruzioni per l uso Consultar las instrucciones de uso Catalog number Num ro catalogue Bes...

Page 9: ...r N sse sch tzen Tenere al riparo dall umidit Fragile Fragile Fr gil Zerbrechlich Fragile This side up Haut Este lado hacia arriba Diese Seite nach oben Alto Date of manufacture Date de production Fec...

Page 10: ...10 IDEXX VetLab UA Analyzer Operator s Guide...

Page 11: ...X VetLab UA Calibration Strips are designed for veterinary use only IDEXX VetLab Station Connectivity The IDEXX VetLab UA Analyzer is part of the IDEXX VetLab suite of analyzers all of which connect t...

Page 12: ...ous wavelengths Reading is done electro optically as follows LED Test pad Test strip Phototransistor A D converter analog to digital Microprocessor Semi quantitative result Cover Press here to open th...

Page 13: ...into an outlet NOTE Use only the power adapter supplied which is marked as supplying 7 5 V DC 3000 mA The cords from other power adapters may t in the power socket of the UA analyzer but may have the...

Page 14: ...he Start button to save your changes and return to the ready to measure mode For more information on setting the language date and time see pages 22 23 3 Before using the UA analyzer it must be calibr...

Page 15: ...tion appears on the display menu See the Troubleshooting section on pages 29 34 for more information IMPORTANT Always wear protective gloves when handling and disposing of veterinary samples Dispose o...

Page 16: ...ted automatically 8 Remove and dispose of the test strip Wipe any urine residue from the tray with a lint free wipe Figure 10 For more information about cleaning the tray see Cleaning the Test Strip T...

Page 17: ...methodology PRO protein neg TR 30 mg dL 100 mg dL 500 mg dL neg TR 0 3 g L 1 0 g L 5 0 g L neg TR 1 2 3 GLU glucose neg 50 mg dL 100 mg dL 300 mg dL 1000 mg dL neg 3 mmol L 6 mmol L 17 mmol L 56 mmol...

Page 18: ...nter the following information on the patient report printout Figure 12 or on the integrated diagnostic report generated by the IDEXX VetLab Station t PMMFDUJPO FUIPE PMM FUI t 7PMVNF 7PM t PMPS t MBS...

Page 19: ...button to return to the Main menu and then clear the memory To Clear Memory The Clear Memory option lets you clear all results from memory Press 1 Menu Press 2 Memory Press 3 Clear and then press Yes...

Page 20: ...options let you specify whether the UA analyzer is operating in a stand alone mode or is connected to other analyzers Press 1 Menu Press 2 Setup 1 Press 3 Modes Press 4 Interf Press an interface opti...

Page 21: ...y or SI arbitrary To Select a Single Unit Option Press 1 Menu Press 2 Setup 1 Press 3 Strip Press 4 Units Press 5 Single Press 6 Conv conventional units mg dL SI SI units or Arbitr arbitrary units 1 2...

Page 22: ...6 No key until you nd the rst parameter that you want to display in the patient report printout Once the desired parameter is found press the Yes key to select the parameter Continue this process unt...

Page 23: ...the t 12 24 button to change the time format Press the t Set button to access keys for changing the actual date and time Press the key to move the cursor from minutes to hours and to year day and mont...

Page 24: ...24 IDEXX VetLab UA Analyzer Operator s Guide...

Page 25: ...rside of the test strip tray can be removed with a soft brush Dry the tray with a lint free wipe IMPORTANT Be careful not to damage the gray reference pad Figure 13 during cleaning and ensure that it...

Page 26: ...ct with urine IMPORTANT Only IDEXX VetLab UA Calibration Strips can be used with the IDEXX VetLab UA Analyzer 4 Press the Start button A beep sounds The tray advances slightly the retaining bar closes...

Page 27: ...mmonia based cleaners ink markers sprays disinfectants containing volatile liquids except isopropyl alcohol insecticides polish or room freshener Care should be taken not to spill any samples chemical...

Page 28: ...28 IDEXX VetLab UA Analyzer Operator s Guide...

Page 29: ...strip tray is soiled or damaged Turn off the analyzer remove the test strip tray carefully clean and dry the tray for instructions see Cleaning the Test Strip Tray on page 25 and then recalibrate the...

Page 30: ...pads on the test strip were not completely moistened with urine Repeat the test using a new IDEXX UA strip Ensure that all of the test pads are completely moistened with urine For more information se...

Page 31: ...ay so that the edge of the tray is ush with the edge of the analyzer Power on the analyzer and wait for the self check process to complete E8 Tray Position Error E9 Wrong Tray Error The test strip tra...

Page 32: ...te E14 Interface Error Data did not transfer to the IDEXX VetLab Station Verify that the analyzer interface option is set to bidirectional mode For more information see Modifying the Interface Options...

Page 33: ...o obtain a relacement test strip tray E16 Reference Pad Error Top Close Printer Cover The printer cover is open Results did not print Close the printer cover and ensure that it locks properly To retri...

Page 34: ...ew IDEXX UA Strip For more information see Reading IDEXX UA Strips on page 15 Strong urine color The analyzer uses the compensation pad on the IDEXX UA Strip to compensate for the strong color Since t...

Page 35: ...ximum 15 W Standby 1 3 W System description Type re ectance photometer Light source six LEDs Wavelengths 565 nm green 3x 610 nm orange 3x Reader head 1 head with six LEDs Measuring cycle approximately...

Page 36: ...I protection The analyzer is classi ed as Category II for over voltage and Degree 2 for pollution according to IEC 664 Opening covers or removing parts of the analyzer except where this can be achieve...

Page 37: ...8 www idexx fr Germany IDEXX GmbH M rikestra e 28 3 D 71636 Ludwigsburg Germany Toll Free Technical Support 00800 1234 3399 Fax 0800 6645627 www idexx de Italy IDEXX Laboratories Italia S r l Via Cano...

Page 38: ...ington ON L7L 4X6 Canada Toll Free Technical Support 1 800 248 2483 Telephone 1 905 602 9499 Toll Free Fax 1 800 248 3010 Fax 1 905 602 6640 idexx com Japan IDEXX Laboratories KK 3 3 7 Kitano Mitaka s...

Reviews: