3
4
61-486
Volt Sense
Range
Min Max
Press 2 Sec
PEAK HOLD
Smart
HOLD
A
Auto
V
LoZ
OFF
MAX
10A
FUSED
MAX
600mA
FUSED
600V CAT IV
1000V CAT III
mA
COM
V
Ω
TRUE RMS
Hz
Hz
V
ac+dc
ac+dc
ac+dc
ac+dc
V
mV
mA
A
Ω
3
• Use appropriate personal protective equipment such as face
shields, insulating gloves, insulating boots, and/or insulating mats.
• Before each use:
- Perform a continuity test by touching the test leads together to
verify the functionality of the battery and test leads.
- Use the 3 Point Safety Method. (1) Verify meter operation by
measuring a known voltage. (2) Apply meter to circuit under
test. (3) Return to the known live voltage again to ensure
proper operation.
• Always work with a partner.
Symbols as marked on the meter and instruction manual
Unsafe Voltage
To alert you to the presence of a potentially hazardous
voltage, when the Tester detects a voltage
≥
30 V or a
voltage overload (OL) in V, mV. The “ “ symbol is
displayed and High voltage indicator is activated.
Meter Description
Front Panel Illustration
1. LCD display : 6000 counts
2. Push-buttons.
3. Rotary switch to turn the Power On / Off and select the
function.
4. Input Terminal for A.
5. Input Terminal for V,
Ω
, , Hz, °C/F° functions.
6. Common (Ground reference) Input Terminal.
7. Input Terminal for mA.
1 Display
2 Buttons
3 Rotary
Switch
5 V,
Ω
,
,
C°/F°
Input
6 Common
Input
7 mA Input
4 A Input
Risk of Danger. Important Information. See Manual.
Hazardous voltage. Risk of electrical shock.
Application around and removal from Hazardous Live
conductors is permitted.
AC (Alternating Current)
DC (Direct Current)
Earth Ground
CAT II
Measurement Category II applies to test and measuring cir-
cuits connected directly to utilization points (socket outlets
and similar points) of the low-voltage MAINS installation
CAT III
Measurement Category III applies to measuring circuits
connected to the distribution part of the building’s low-
voltage MAINS installation
CAT IV
Measurement Category IV applies to test and measuring
circuits connected at the source of the building’s low-
voltage MAINS installation
~
...