1.2
Specifications
Supply voltage
12 … 28 V (dc), 2 W protected
against power reversal and tran-
sient over-voltages
Connection
2-pin pluggable terminal block
(mating connector included)
Permissible temperature
Operating
0 … +40 °C
Storage
–40 … +70 °C
Relative humidity
0 to 95%, non-condensing
Use
Operating
altitude up to 2500 m (8200 ft.)
Storage
altitude up to 12500 m (41000 ft.)
Housing
1/8-DIN panel-mount enclosure
(aluminum extrusion)
Connections
Gauge
9-pin D-sub female (mating
connector provided as part of the
gauge cable)
Analog output and serial
communications interface
9-pin D-sub male
Relay outputs
6-pin pluggable terminal block
(mating connector included)
Measurement range
1.3×10
-4
… 1333 mbar
1×10
-4
… 1000 Torr
1.3×10
-2
Pa … 133 kPa
Display rate
0.5 s
Units
mbar, Torr or Pa
Display
bright OLED, 4 digits
1100 Torr … 1000 Torr
4 digits
999 Torr … 10.0 mTorr
3 digits
9.9 mTorr … 1.0 mTorr
2 digits
0.9 mTorr … 0.1 mTorr
1 digit