V100 (2019-11)
Doc. Ref. XGC-320-100119
43
Notes:
1) OP = Overpressure indication; display will read
"overpressure".
Examples:
1) Gas used is nitrogen. Display shows pressure measurement
of 13.3 mbar. True pressure of nitrogen is 13.3 mbar.
2) Gas used is argon. Display shows pressure measurement of
11.7 mbar. True pressure of argon is 133 mbar.
3) Gas used is CO
2
. Display shows pressure measurement of
.0731 mbar. True pressure of CO
2
is .0666 mbar.