V100 (2019-11)
Doc. Ref. XGC-320-100119
57
Note - when using the units of pascals and LOG 1-8 is selected,
the same equation of P = 10
(V - 5)
listed above applies. This re-
sults in a log-linear analog output range of about 3.00 V (dc) at
.01 pascals (Pa) and 10.12 V (dc) at 133 kPa.
Chart of the calculated pressures using the formulas and data for
the log-linear output signal for nitrogen from the previous page.
7.3
Log 0-7; Log-Linear Analog Output Equation
& Table - Torr
Log-Linear 0 to 7 V analog output for selected gases - Engineer-
ing units in Torr (see following table):