Page 21
ACV High Frequency Calibration
Calibrating the ACV Frequency Response
1. 400V Range
Set the calibrator to generate 200.00V / 10KHz
±
0.07% at least, adjust VC3 to
have a reading of 200.50
±
20 count .
2. 40V Range
Set the calibrator to generate 20.000V /10KHz
±
0.07% at least, adjust VC2 to
have a reading of 20.050
±
20 count .
3. 4V Range
Set the calibrator to generate 2.0000V /10KHz
±
0.07% at least, adjust VC1 to
have a reading of 2.0050
±
20 count .
Form Number TM61633-5
Rev 5 Dec 2006