Page 20
Figure 4
Calibration:
Calibration of the meter is recommended once a year to ensure performance according to the published
specifications.
Calibration is performed with the use of “CalWin 300” Software as illustrated above.
(This software can be obtained by contacting customer service at Ideal Industries, Inc.)
Do not perform this step if software is already loaded
The software consists of two files titled “Setup.exe” and “Data001”. After you execute the setup.exe program in
the Windows (95 or 98) environment, you will have CalWin300 option in the PROGRAM set.
Execute “CalWin300” and you should get the screen shown in first Figure 3.
Description of CalWin300 window shown in Figure 4.
1
. Main Display:
Displays the main reading of the Meter.
2
. Connection Indicator:
“Connection” shows that the meter is connected to your
Computer and the meter and computer are communicating.
“Disconnection” shows that the meter and computer are not
connected or aren’t communicating.
3
. Communication Port:
Selects port available to connect to meter.
Selectable ports are COM 1 or COM 2.
4.
Calibration Function Indicator and Selector:
Select item to be calibrated and select the Calibration Function desert.
5.
Fine Frequency Selector:
Selects the Frequency that’s connected the meter, to select 50Hz or 60Hz.
Note:
Before the Calibration Function is selected, the function selected on the Meter has to be
confirmed, to prevent functional error from occurring.
Form Number TM61633-5