______________________________________________________________________
Co
Brite
User Manual
34
Syntax
Read
Write
User
access
level
Paramete
r format
Description
[:SYStem:]ALARm?
R
0
xxxxx
Query alarm status of device,
see Alarm code definition for
details
[:SYStem:]STATus?
R
0
“INIT”;”FAU
LT”,”READ
Y”;”OVERL
OAD”
Query current state of OSA; see
Alarm state for match
[:SYSTem:]ERRor[:NEX
T]?
R
-
Queries data in an error queue
and deletes it from the queue.
[:SYSTem:COMMunicate
:]LOCKout/?
RW
1
“0” or “1”
Locks other sessions to perform
write commands on the unit. It is
automatically released if the
active session closes
[:SYSTem:COMMunicate
:]ParameterREFresh?
R
0
<INT>
This query allows detecting if
any change to the 34eparately
configuration was done. Each
time the counter increases, the
configuration was changed. This
is useful in multi-user
environments to detect if a
parallel session has changed
settings on the unit
[:SYStem:]IPConfCHan
ged?
R
0
0|1
Check if IP configuration of the
Ethernet Interface was changed
but no reboot was triggered yet
to activate the new settings. This
can be used to determine if the
actual settings differ from the
settings that will be active upon
restart.
[:STATus:OPERation:]
EVENt?
R
0
-
Read status of instrument if
scan has been completed (1) or
not (0). Each read of
“EVEN?”
will clear the status for each
remote session 34eparately.
This can be used to query if
previous scan has been
completed.
[:]ABORt
W
0
-
Aborts all currently executed
pending commands as fast as
possible. Query *OPC? To
determine status once all
pending commands are aborted
[:BOOT:]BootLoaDerRE
Set
W
0
-
Resets upload procedure for
firmware upgrade. All data
uploaded already is erased.
New process can be restarted
[:BOOT:]
BootLoaDerAPPend
W
0
binary
send portions of the binary
Firmware Image, each Byte is
coded into 2 Characters 1..F,
send max. 100 Bytes per
command