User Manual
32
:
Usage of Colon see 7.3.4
<wsp>
Denominates a white space character
<INT>
Integer value
<REAL>
Real value
7.6
General SCPI Commands
Syntax
Example
Description
*idn?
*idn?;
COBRITE CBDX-NC-
NN-NN-NN-FA, SN
19160001, F/W Ver
1.0.0(101), HW Ver
1.00;
Query System Type and Software version. The
CoBrite DX is identified by CBDX. The
CoBriteDX2 by CBDX2. The remaining string is
determined by the laser configuration of the actual
device. It matches the part number that is printed
on the label on the unit.
*opc?
*opc?;
1;
”
Query if all pending commands were executed.
Note that this will NOT indicate if physical tuning of
laser ports has been completed. Use
busy?
Query for this.
*rst
*rst;
Reset Controller; will warm start the instrument. All
connections/sessions will be closed. Requires
user level 1
*WAIt
*WAI;
;
Unit waits with response until *opc? is 1.
Eliminates need for polling loop of *opc? on
remote side
*CLS
*CLS;
;
Clears all status/alarm registers of the unit. This
command is used to clear latched alarm registers.
7.7
Commands on system level
Syntax
Read
Write
User
access
level
Paramete
r format
Description
[:SYStem:]ECHO/?/<ws
p><P>
RW
0
“0” or “1”
Sets or query echo command
sent status, this setting applies
to current session only
[:SYStem:]DEFAULT
W
-
Resets user settings such as
laser settings to factory default.
This affects all settings including
trigger configuration except
Remote interface settings.
[:SYStem:]STArtDEFau
lt /?<wsp>/<P>
RW
1
“0” or “1”
Queries/Sets of unit starts with
last settings set before a reboot
or with default of device. This
affects all settings except
Remote interface settings and
Laser Port on/off status.
0 = Save each change of setting
and restart with current config
upon next reboot
1 = Always start with unit factory
defaults