11
•
Press F1 during 2 seconds to leave the average mode
Note: the effort must be an actual value (neither "–MIN–" nor "–MAX–») otherwise, the display
shows the error message “!”
8.2. Record
To record the data at the end of a measurement process
•
Press key " " during the measurement process, display will show "REC"
•
Conditions : Wire rope selected / Batteries OK
•
In average mode, the value recorded is the average value and not the current effort
•
To read the values recorded, enter the menu : records : read
Note: the effort must be an actual value (neither "–MIN–" nor "–MAX–») otherwise, the display
shows the error message “!”
8.3. Custom calibration
8.3.1 Create
This process (carried out by user) allows to correct the “raw effort” signal provided by the
sensor.
The special calibration is valid for only one type of cable at a single value of effort.
This operation must only be performed using the unit daN both on the tension meter
and on the dynamometer (see § 7).
To perform a custom calibration, place an online dynamometer in the installation and apply the
desired tension in the cable.
•
Press F2 and select the cable -- Special -- (first line of the list)
•
Select a blank line “Spc xx /”. Press F2
•
Select "Calibrate" with key
to create a new calibration
o
"Use" see § 8.3.2
o
"Delete" to delete an existing custom calibration
•
Install the tension meter on the cable (see § 6.1)
•
Your display shows the “raw effort” (see § 4.1).
Make sure the raw effort never is above 1.500 (daN). If it is the case, take the
tension meter off the cable.
o
To have enough signal, the raw effort must be higher than 150
•
Press the key
•
Adjust the value shown by the display at the same value as shown by the dynamometer
using keys T and
•
Save the value with
•
End process with F1
•
The display shows the current effort with, at the bottom left of the screen, the reference
of the custom calibration