27
I-7011, I-7018 User Manual
Rev:B1.0
2.6 $AA1
Description
: Perform Zero Calibration
Syntax
:
$AA1[CHK](cr)
$
delimiter character
AA
address of setting module (00 to FF)
1
command for performing zero calibration
Response
:
Valid Command :
!AA[CHK](cr)
Invalid Command :
?AA[CHK](cr)
Syntax error or communication error may get no re-
sponse.
!
delimiter for valid command
?
delimiter for invalid command or the calibration is not
enabled
AA
address of response module(00 to FF)
Example
:
Command : $011
Receive : !01
Perform address 01 zero calibration, return success.
Command : $021
Receive : ?02
Perform address 02 zero calibration, return the calibration is
not enabled before perform calibration command.
Related Command
:
Sec.2.5
$AA0,
Sec.2.20
~AAEV
Related Topics
:
Sec.1.9
Calibration