19. ELECTROMAGNETIC DECLARATION
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because
it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING
Use of accessories other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Icare PRO (TA03) is class BF-type equipment and needs special precautions regarding EMC and
needs to be installed and put into service according to EMC information provided in user manual.
GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC EMISSIONS
Icare PRO (TA03) is intended for use in the electromagnetic environment specified below.
The user of the Icare PRO (TA03) should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare PRO (TA03) is battery operated and uses RF
energy only for its internal function. Therefore,
its RF emissions are low and are not likely to
cause any interference in nearby equipment.
RF emissions CISPR 11
Class B
Icare PRO (TA03) is suitable for use in
all establishments, including domestic
establishments and those directly connected to
public low-voltage power supply network that
supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Power level of Icare PRO (TA03) is below
standard requirement
Voltage fluctuations
flickering emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY
Icare PRO (TA03) is intended for use in the electromagnetic enviroment specified below. The customers
or users of Icare PRO (TA03) should assurre that it is used in such enviroment.
Immunity test
IEC 60601 Test level Compliance
level
Electromagnetic environment-
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
Transients/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
NOT
APPLICABLE
Icare PRO (TA03) tonometer is not
operational when connected to an
external power supply
Surge
IEC 61000-4-5
±1 kV for line(s) to
line(s)
±2 kV for line(s) to
earth
NOT
APPLICABLE
Icare PRO (TA03) tonometer is not
operational when connected to an
external power supply
Summary of Contents for Pro TA03
Page 2: ...3 4 5 6 3 7 mm 1 8 9 32 90 Icare tonometer PRO manual 1 2...
Page 3: ...8 11 12 13 14 15 9 10 7...
Page 17: ......
Page 33: ......
Page 75: ......
Page 89: ......
Page 103: ......
Page 131: ......
Page 174: ......
Page 175: ......
Page 176: ...www icaretonometer com...