Icare HOME Patient Guide

ENGLISH

17

www.icare-usa.com

16

Probe base indicator light is 

solid red and the Measure 

indicator light turns off.

Cause: Tonometer is not held

straight.

Probe base indicator light is 

flashing red and you hear 2 

beeps.

Cause: Tonometer is too far

from or too close to eye or

not held straight.

 

Probe base indicator light is 

flashing red and Repeat light 

is flashing and you hear 2 

beeps.

Cause: Too much variation

between measurements or

eye recognition failed.

 

Probe base indicator light is 

flashing red and Service light 

is flashing and you hear 2 

beeps.

Cause: The probe or probe 

base is dirty or damaged.

 

Battery light is flashing.

Cause: Battery power is low.

 

ERROR MESSAGE

1. 

Make sure you are facing straight

  ahead (i.e., head held at a 90° angle).

2. 

Adjust the tonometer’s tilt angle until

  probe base light turns green.

1. 

Press the Measure button         to clear error.

2. 

Adjust the forehead/cheek supports for

  correct distance. 

3. 

Repeat measurement.

1. 

Press the Measure button         to clear error.

2. 

Make sure the eye recognition sensors or

  temple side of the eye are not covered  

  as covering them causes the eye recognition  

  to fail and the error.

3. 

Repeat measurement.

Contact the health care professional you received 

your tonometer from or a certified service center 

to have the tonometer serviced.

Change the battery.

See pages 18-19 for instructions.

SOLUTION

Summary of Contents for Icare HOME TA022

Page 1: ...1 www icare usa com PATIENT ENGLISH Federal US law restricts this device to sale by or on the order of a physician or properly licensed practitioner Belongs to GUIDE...

Page 2: ...clear layer Forehead cheek supports tonometer s adjustable supports Expected service life expected life time before replacement Electromagnetic immunity tonometer s ability to withstand disturbances...

Page 3: ...have taken the measurements for both eyes The tonometer includes infra red eye recognition sensors to identify which eye right or left you are measuring It is important not to cover these sensors with...

Page 4: ...oor or off center visual fixation Have poor hearing without an aid and or communicate using sign language Use contact lens Have dry eyes Have Keratoconus a corneal disorder Have a congenitally from bi...

Page 5: ...indicator REPEAT indicator DONE indicator SERVICE indicator BATTERY indicator Power button Cheek support position indicator Device label Cheek support adjustment wheel Silicone lid USB cover Battery c...

Page 6: ...es made by the manufacturer Do not touch the probe The shelf life of the probes is 3 years Use a new unused probe after measure ments for both eyes have been taken You should thoroughly wash your hand...

Page 7: ...the probe tip points straight at cornea WARNING Make sure that the probe or any part of the tonometer does not contact your eyes PRECAUTION Do not cover the eye recognition sensors because covering th...

Page 8: ...onometer to measure your eye pressure you can remove the probe by powering off the device and allowing the probe to slide out of the probe base Place the probe back in its container and dispose of it...

Page 9: ...The probe or probe base is dirty or damaged Battery light is flashing Cause Battery power is low ERROR MESSAGE 1 Make sure you are facing straight ahead i e head held at a 90 angle 2 Adjust the tonom...

Page 10: ...d batteries and insert 2 new CR123A lithium batteries Make sure the batteries positive end is facing the battery cover Rechargeable batteries should not be used Slide the battery cover back over the b...

Page 11: ...damp cloth or paper towel Make sure the batteries positive end is facing the battery cover PRECAUTION Do not immerse the tonometer in water or any other liquid PRECAUTION Any maintenance services othe...

Page 12: ...f 5 to 50 mmHg Compared to the Goldmann tonometer less than 5 of measurements fell outside 5 mmHg at each pressure range and less than 1 fell outside 7 5 mmHg at each pressure range The Icare HOME s v...

Page 13: ...mbined with ISO 13485 certified quality system has made us a respected player in our field of expertise The advanced Icare product line offers reliable high pre cision reproducible accuracy in measuri...

Reviews: