Icare HOME Patient Guide

ENGLISH

11

www.icare-usa.com

10

Support position tag

Support position tag

Make sure the forehead and cheek support 

positions match the positions your eye care 

professional wrote on the support position 

tag for you.

The support position tag is found in the Icare 

tonometer carrying case. If needed, turn the 

support wheels to your set support settings.

Press and release the Power button          to

power on the tonometer. All indicator lights 

on the back panel will flash once and you will 

hear a short beep.

The LOAD indicator will continue to flash

indication that the tonomoter is ready to

load the probe.

Gather new probe and unwrap package.

     

WARNING

•  Use only original and certified probes made  

  by the  manufacturer.

•  Do not touch the probe.

•  The shelf life of the probes is 3 years.

•  Use a new, unused probe after measure-   

  ments for both eyes have been taken.

•  You should thoroughly wash your hands with  

  soap and water and dry them with a clean  

  towel prior to handling the probe.

Hold the probe container upright, with the 

container cap on top and uncap the container, 

tilt the container to allow the probe to drop into 

the probe base, then press the Measure button           

briefly (1 second). 

The flash indicator light stops once the probe 

is loaded and the MEASURE indicator starts to 

flash when the tonometer is ready to measure. 

Keep the probe container to reuse when ready 

to dispose of the probe.

     

WARNING

•  Do not touch the probe.

•  Do not point the tonometer down.

•  Do not use the probe if it touches a non-

  sterile surface like table or floor.

PREPARING THE TONOMETER 

FOR USE

1

2

3

4

Cheek support

Forehead support

Summary of Contents for Icare HOME TA022

Page 1: ...1 www icare usa com PATIENT ENGLISH Federal US law restricts this device to sale by or on the order of a physician or properly licensed practitioner Belongs to GUIDE...

Page 2: ...clear layer Forehead cheek supports tonometer s adjustable supports Expected service life expected life time before replacement Electromagnetic immunity tonometer s ability to withstand disturbances...

Page 3: ...have taken the measurements for both eyes The tonometer includes infra red eye recognition sensors to identify which eye right or left you are measuring It is important not to cover these sensors with...

Page 4: ...oor or off center visual fixation Have poor hearing without an aid and or communicate using sign language Use contact lens Have dry eyes Have Keratoconus a corneal disorder Have a congenitally from bi...

Page 5: ...indicator REPEAT indicator DONE indicator SERVICE indicator BATTERY indicator Power button Cheek support position indicator Device label Cheek support adjustment wheel Silicone lid USB cover Battery c...

Page 6: ...es made by the manufacturer Do not touch the probe The shelf life of the probes is 3 years Use a new unused probe after measure ments for both eyes have been taken You should thoroughly wash your hand...

Page 7: ...the probe tip points straight at cornea WARNING Make sure that the probe or any part of the tonometer does not contact your eyes PRECAUTION Do not cover the eye recognition sensors because covering th...

Page 8: ...onometer to measure your eye pressure you can remove the probe by powering off the device and allowing the probe to slide out of the probe base Place the probe back in its container and dispose of it...

Page 9: ...The probe or probe base is dirty or damaged Battery light is flashing Cause Battery power is low ERROR MESSAGE 1 Make sure you are facing straight ahead i e head held at a 90 angle 2 Adjust the tonom...

Page 10: ...d batteries and insert 2 new CR123A lithium batteries Make sure the batteries positive end is facing the battery cover Rechargeable batteries should not be used Slide the battery cover back over the b...

Page 11: ...damp cloth or paper towel Make sure the batteries positive end is facing the battery cover PRECAUTION Do not immerse the tonometer in water or any other liquid PRECAUTION Any maintenance services othe...

Page 12: ...f 5 to 50 mmHg Compared to the Goldmann tonometer less than 5 of measurements fell outside 5 mmHg at each pressure range and less than 1 fell outside 7 5 mmHg at each pressure range The Icare HOME s v...

Page 13: ...mbined with ISO 13485 certified quality system has made us a respected player in our field of expertise The advanced Icare product line offers reliable high pre cision reproducible accuracy in measuri...

Reviews: