8. READING THE MEASUREMENT DATA

The tonometer stores information on every complete measurement sequence of six 

measurements. The stored information includes the calculated final eye pressure 

reading in mmHg, time and date of the measurement, identification of the eye (right 

or left) and the quality level of the measurement.

9. TROUBLESHOOTING

The tonometer automatically monitors and controls the measurement position and 

speed of the probe during the measurements, and indicates errors with sounds 

and lights. The following table instructs you in error situations and explains what 

the different lights and sounds mean. The indicator lights are also presented in the 

figure below the table.

Error light

Error 

sound

Reason

Action

Battery

No.

Battery is soon 

empty.

Prepare to change batteries.

Battery light is flashing

No.

Battery is empty.

Change batteries.

Probe base light is solid red.

No.

Too much vertical 

tilt.

Press the measurement button again to clear the 

error message.

Position the tonometer horizontally so that the 

probe base light is green.

Probe base light is flashing red 

and Measure light turns off.

Two long 

beeps.

a) Probe is too far 

from or too near 

the eye.

b) Probe 

movement was not 

perpendicular to 

the cornea.

Press the measurement button again to clear the 

error message.

a) Set the correct distance 4-8mm (5/32-5/16”) 

between the probe tip and the center of the cornea.

b) Set the probe perpendicular to the center of the 

cornea.

Repeat light is flashing and 

Probe base light is flashing red.

Two long 

error 

beeps.

a) Too much IOP 

deviation during 

the measurement, 

because the user 

did not keep the 

tonometer stable.

b) Eye was not 

recognized. 

Press the measurement button again to clear the 

error message.

a) Repeat the measurement

b) Do not move the tonometer during the 

measurements, remove your hand or fingers from 

the infrared transmitters and sensor, move the 

patient’s hair away from his/her temple side of the 

eye.

Service light is flashing and 

Probe base is flashing red.

Two long 

beeps.

Incorrect or dirty 

probe or probe 

base. 

Contact the seller to arrange sending the device for 

service.

DONE

SERVICE

BATTERY

BATTERY

REPEAT

SERVICE

Icare HOME tonometer Instruction manual for health care professionals

Icare HOME tonometer Instruction manual for health care professionals

English

English

More information about 

Icare LINK software

http://www.icaretonometer.com/products/icare-link/

Uploading is easy:

1.  Start Icare LINK software in your PC.

2.  Connect the tonometer to the PC using the USB 

cable. The Load and Measure lights will flash. If 

no lights flash or the Service and Battery lights 

flash, reconnect the USB cable.

3.  The internal clock of the tonometer is 

automatically updated to the PC’s time by the 

Icare LINK software at this point.

4.  Copy the data to a selected patient in the Icare 

LINK software.

LINK

Connected

Summary of Contents for Icare HOME TA022

Page 1: ...HOME INSTRUCTION MANUAL FOR HEALTH CARE PROFESSIONALS ENGLISH...

Page 2: ...modification of this equipment is allowed WARNING Use only the original and certified probes made by the manufacturer The probes are for single use single pair of measurement sequences only Use probes...

Page 3: ...arby devices for example sensitive sensors If you do not use the tonometer for a long time remove the batteries as they may leak Be sure to dispose of the single use probes properly for example in a c...

Page 4: ...rly for damage to the case If you suspect that there is something wrong with the tonometer contact the dealer who sold the tonometer to you The package contains Icare HOME tonometer 10 sterilized sing...

Page 5: ...re turned on briefly Following a brief pause the Load light flashes on the back panel to remind the user to load the single use probe into the tonometer prior to measurement 6 3 LOADING THE PROBE WARN...

Page 6: ...ach of the six measurements to obtain the final IOP reading see figure left Series mode press the measurement button constantly Single mode press the measurement button briefly 1 second for each of th...

Page 7: ...forehead and cheek supports with adjustment wheels 7 3 AUTOMATIC EYE RECOGNITION NOTE Do not cover the eye recognition transmitters or sensor during the measurement for example with your fingers Keep...

Page 8: ...sition Incorrect measurement position Incorrect head and eye position Incorrect measurement position 5 Press the measurement button Single mode Press the button briefly 1 second and you hear a short b...

Page 9: ...message a Set the correct distance 4 8mm 5 32 5 16 between the probe tip and the center of the cornea b Set the probe perpendicular to the center of the cornea Repeat light is flashing and Probe base...

Page 10: ...At the same time the probe base light flashes red and you hear two short beeps The reason is either too much deviation in your measurement or the automatic eye recognition system could not recognize...

Page 11: ...the tonometer Explain that the measurement button must be depressed to obtain the sequence of 6 measurements until a long beep is heard and the green Done light is illuminated on the back panel The pr...

Page 12: ...certified prior to performing self tonometry STEP 1 Instruct the patient 1 Show and explain the tonometer s user interface including icons and status indicators Off no lights no signals Press power b...

Page 13: ...strate measurements from your HCP s own eye 1 Position the tonometer on your HCP own eye as instructed above 2 Ask the patient to observe and learn STEP 3 Supervised patient use of the icare home 1 Lo...

Page 14: ...strictions for professional use include Medivac vehicles or similar where vibration or noise levels are so high that the user cannot hear error signals Environmental restrictions for lay operators pat...

Page 15: ...in all establishments including domestic establishments and those directly connected to public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufact...

Page 16: ...he location in which the Icare HOME is used exceeds the applicable RF compliance level above the Icare HOME should be observed to verify normal operation If abnormal performance is observed additional...

Page 17: ...www icaretonometer com 32 Icare HOME www icaretonometer com HOME Icare HOME tonometer Instruction manual for health care professionals English...

Reviews: