8. READING THE MEASUREMENT DATA
The tonometer stores information on every complete measurement sequence of six
measurements. The stored information includes the calculated final eye pressure
reading in mmHg, time and date of the measurement, identification of the eye (right
or left) and the quality level of the measurement.
9. TROUBLESHOOTING
The tonometer automatically monitors and controls the measurement position and
speed of the probe during the measurements, and indicates errors with sounds
and lights. The following table instructs you in error situations and explains what
the different lights and sounds mean. The indicator lights are also presented in the
figure below the table.
Error light
Error
sound
Reason
Action
Battery
No.
Battery is soon
empty.
Prepare to change batteries.
Battery light is flashing
No.
Battery is empty.
Change batteries.
Probe base light is solid red.
No.
Too much vertical
tilt.
Press the measurement button again to clear the
error message.
Position the tonometer horizontally so that the
probe base light is green.
Probe base light is flashing red
and Measure light turns off.
Two long
beeps.
a) Probe is too far
from or too near
the eye.
b) Probe
movement was not
perpendicular to
the cornea.
Press the measurement button again to clear the
error message.
a) Set the correct distance 4-8mm (5/32-5/16”)
between the probe tip and the center of the cornea.
b) Set the probe perpendicular to the center of the
cornea.
Repeat light is flashing and
Probe base light is flashing red.
Two long
error
beeps.
a) Too much IOP
deviation during
the measurement,
because the user
did not keep the
tonometer stable.
b) Eye was not
recognized.
Press the measurement button again to clear the
error message.
a) Repeat the measurement
b) Do not move the tonometer during the
measurements, remove your hand or fingers from
the infrared transmitters and sensor, move the
patient’s hair away from his/her temple side of the
eye.
Service light is flashing and
Probe base is flashing red.
Two long
beeps.
Incorrect or dirty
probe or probe
base.
Contact the seller to arrange sending the device for
service.
DONE
SERVICE
BATTERY
BATTERY
REPEAT
SERVICE
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
English
English
More information about
Icare LINK software
http://www.icaretonometer.com/products/icare-link/
Uploading is easy:
1. Start Icare LINK software in your PC.
2. Connect the tonometer to the PC using the USB
cable. The Load and Measure lights will flash. If
no lights flash or the Service and Battery lights
flash, reconnect the USB cable.
3. The internal clock of the tonometer is
automatically updated to the PC’s time by the
Icare LINK software at this point.
4. Copy the data to a selected patient in the Icare
LINK software.