www.icaretonometer.com
www.icaretonometer.com
26
27
Icare HOME
Icare HOME
14. TECHNICAL AND PERFORMANCE DATA
Type TA022
Dimensions: approximately 11cm x 8cm x 3cm.
Weight:
approximately 150g.
Power supply:
2 x CR123 non-rechargeable batteries (make sure you use batteries with
built-in PTC protection, for example Energizer Lithium Photo 123 3V CR123A).
Measurement range:
5-50 mmHg.
Accuracy
(95% tolerance interval relatively to manometry): ±1,2mmHg (<20 mmHg) and
±2.2 mmHg (≥20 mmHg).
Repeatability
(coefficient of variation): < 8%.
The serial number is located on the inside of the battery compartment cover.
The lot number of the probes is on the side of the probe box and the blister packing.
There are no electrical connections from the tonometer to the patient.
The tonometer has BF-type electric shock protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30% to 90%
Atmospheric pressure: 800hPa – 1060hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10% to 95%
Atmospheric pressure: 700hPa – 1060hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10% to 95%
Atmospheric pressure: 500hPa – 1060hPa
Environmental restrictions for professional use include:
• Medivac vehicles or similar where vibration or noise levels are so high
that the user cannot hear error signals.
Environmental restrictions for lay operators (patients):
• Environments where noise is so high that the user cannot hear the error
signals.
Mode of operation:
continuous
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
English
English
12. CLEANING AND DISINFECTION
NOTE!
Do not carry out any other service procedures. Leave all other service and repairs to the
manufacturer or a certified service center.
You must clean the forehead and cheek supports for each new patient. Use a wipe
dampened with a 70% isopropyl alcohol solution. Do not immerse the tonometer in
water or other liquid. The tonometer must not be immersed or cleaned using too
much water.
13. ACCESSORIES
Part number
Product Description
Weight Dimensions
540
Probe base
4g
7 mm x 38 mm
TA022-001
Probe base collar
2g
20 mm x 15 mm
560
Wrist strap
3g
270 mm x 10 mm x 10 mm
TA022-044
Carrying case
210g
270 mm x 135 mm x 60 mm
7179
Battery cover
3g
26 mm x 23 mm x 7 mm
TA022-035
Patient guide
33g
210 mm x 90 mm x 2 mm
571
Battery 3 V, CR123A 17g
17 mm x 35 mm
TA022-037
Support position tags 40g
70 mm x 41 mm x 13 mm
575
USB cable
23g 1m