14.  TECHNICAL AND PERFORMANCE DATA

Type TA022 

Dimensions: approximately 11cm x 8cm x 3cm. 

Weight:

 approximately 150g. 

Power supply:

 2 x CR123 non-rechargeable batteries (make sure you use batteries with 

built-in PTC protection, for example Energizer Lithium Photo 123 3V CR123A). 

Measurement range:

 5-50 mmHg. 

Accuracy

 (95% tolerance interval relatively to manometry): ±1,2mmHg (<20 mmHg) and 

±2.2 mmHg (≥20 mmHg). 

Repeatability

 (coefficient of variation): < 8%. 

The serial number is located on the inside of the battery compartment cover. 

The lot number of the probes is on the side of the probe box and the blister packing.

There are no electrical connections from the tonometer to the patient. 

The tonometer has BF-type electric shock protection. 

Operation environment:

Temperature: +10 °C to +35 °C 

Relative humidity: 30% to 90% 

Atmospheric pressure: 800hPa – 1060hPa 

Storage environment:

Temperature: -10 °C to +55 °C

Relative humidity: 10% to 95% 

Atmospheric pressure: 700hPa – 1060hPa

Transport environment:

 

Temperature: -40 °C to +70 °C

Relative humidity: 10% to 95% 

Atmospheric pressure: 500hPa – 1060hPa 

Environmental restrictions for professional use include:

•  Medivac vehicles or similar where vibration or noise levels are so high 

that the user cannot hear error signals.

Environmental restrictions for lay operators (patients):

•  Environments where noise is so high that the user cannot hear the error 

signals.

Mode of operation: 

continuous

Icare HOME tonometer Instruction manual for health care professionals

Icare HOME tonometer Instruction manual for health care professionals

English

English

12.  CLEANING AND DISINFECTION

 NOTE!

Do not carry out any other service procedures. Leave all other service and repairs to the 

manufacturer or a certified service center.

You must clean the forehead and cheek supports for each new patient. Use a wipe 

dampened with a 70% isopropyl alcohol solution. Do not immerse the tonometer in 

water or other liquid. The tonometer must not be immersed or cleaned using too 

much water.

13.  ACCESSORIES

Part number

Product Description

Weight Dimensions

540  

Probe base 

 4g 

7 mm x 38 mm

TA022-001 

Probe base collar 

 2g 

20 mm x 15 mm

560  

Wrist strap 

 3g 

270 mm x 10 mm x 10 mm

TA022-044   

Carrying case 

 210g 

270 mm x 135 mm x 60 mm

7179  

Battery cover 

 3g 

26 mm x 23 mm x 7 mm

TA022-035 

Patient guide 

 33g 

210 mm x 90 mm x 2 mm

571  

Battery 3 V, CR123A   17g 

17 mm x 35 mm

TA022-037 

Support position tags   40g 

70 mm x 41 mm x 13 mm

575  

USB cable  

23g 1m

Summary of Contents for Icare HOME TA022

Page 1: ...HOME INSTRUCTION MANUAL FOR HEALTH CARE PROFESSIONALS ENGLISH...

Page 2: ...modification of this equipment is allowed WARNING Use only the original and certified probes made by the manufacturer The probes are for single use single pair of measurement sequences only Use probes...

Page 3: ...arby devices for example sensitive sensors If you do not use the tonometer for a long time remove the batteries as they may leak Be sure to dispose of the single use probes properly for example in a c...

Page 4: ...rly for damage to the case If you suspect that there is something wrong with the tonometer contact the dealer who sold the tonometer to you The package contains Icare HOME tonometer 10 sterilized sing...

Page 5: ...re turned on briefly Following a brief pause the Load light flashes on the back panel to remind the user to load the single use probe into the tonometer prior to measurement 6 3 LOADING THE PROBE WARN...

Page 6: ...ach of the six measurements to obtain the final IOP reading see figure left Series mode press the measurement button constantly Single mode press the measurement button briefly 1 second for each of th...

Page 7: ...forehead and cheek supports with adjustment wheels 7 3 AUTOMATIC EYE RECOGNITION NOTE Do not cover the eye recognition transmitters or sensor during the measurement for example with your fingers Keep...

Page 8: ...sition Incorrect measurement position Incorrect head and eye position Incorrect measurement position 5 Press the measurement button Single mode Press the button briefly 1 second and you hear a short b...

Page 9: ...message a Set the correct distance 4 8mm 5 32 5 16 between the probe tip and the center of the cornea b Set the probe perpendicular to the center of the cornea Repeat light is flashing and Probe base...

Page 10: ...At the same time the probe base light flashes red and you hear two short beeps The reason is either too much deviation in your measurement or the automatic eye recognition system could not recognize...

Page 11: ...the tonometer Explain that the measurement button must be depressed to obtain the sequence of 6 measurements until a long beep is heard and the green Done light is illuminated on the back panel The pr...

Page 12: ...certified prior to performing self tonometry STEP 1 Instruct the patient 1 Show and explain the tonometer s user interface including icons and status indicators Off no lights no signals Press power b...

Page 13: ...strate measurements from your HCP s own eye 1 Position the tonometer on your HCP own eye as instructed above 2 Ask the patient to observe and learn STEP 3 Supervised patient use of the icare home 1 Lo...

Page 14: ...strictions for professional use include Medivac vehicles or similar where vibration or noise levels are so high that the user cannot hear error signals Environmental restrictions for lay operators pat...

Page 15: ...in all establishments including domestic establishments and those directly connected to public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufact...

Page 16: ...he location in which the Icare HOME is used exceeds the applicable RF compliance level above the Icare HOME should be observed to verify normal operation If abnormal performance is observed additional...

Page 17: ...www icaretonometer com 32 Icare HOME www icaretonometer com HOME Icare HOME tonometer Instruction manual for health care professionals English...

Reviews: