TECHNICAL INFORMATION

The serial number is on the inside of the battery compartment cover. There are no electrical connections from the tonometer to the patient. The device has  

BF-type electric shock protection. The single use probe and the forehead support of the device are considered as applied parts.

IT-NETWORK SPECIFICATIONS

 

WARNING

Connection of the ic200 tonometer to IT networks including other equipment could result in previously unidentified risks to patients, operators, or third parties.

 

WARNING

The responsible organization should identify, analyze, evaluate, and control any additional risks resulting from the ic200 tonometer connected to IT networks 

including other equipment.

 

PRECAUTION

Changes to the IT network could introduce new risks requiring additional analysis by the responsible organization. The changes include:

•  changes in the IT-network configuration

•  connection of additional items to the IT-network

•  disconnecting items from the IT-network

•  update or upgrade of equipment connected to the IT-network

To transfer the measurement data from the ic200 tonometer to a host device, the tonometer must be connected to responsible organizations IT Network over 

Bluetooth. The ic200 tonometer is designed to work stand-alone without the Bluetooth connection. The ic200 tonometer is designed such that network failures 

will not prevent the ic200 tonometer from working normally.

Required characteristics of the IT-network:

Printer: Bluetooth® Classic, ESC/POS communication protocol.

Computer: Bluetooth® 4.0 (or greater) Low Energy support.

Connection is secured by link authentication.

Intended Information Flow:

Measurement data is collected by the ic200 Tonometer. This data is sent via Bluetooth connection either to a wireless printer (Bluetooth Classic), or to the 

computer (Bluetooth Low Energy, BLE) which has the Icare Export application installed. The Icare Export will transfer the data into the Icare CLINIC software. 

Operators can then access the data by using the Icare CLINIC software with a web browser in an internet connected device.

Potential Hazardous situations resulting from the failure of the IT-network:

•  If the Bluetooth connection is lost during data transfer, no data is lost from the device. The measurement data can still be found from the device history and 

transferred once the connection is re-established.

•  Failure or misconfiguration of the IT-network may result in data not being transferred causing inconvenience to the operators.

ACCESSORIES

SKU

PRODUCT DESCRIPTION

WEIGHT

DIMENSIONS

104

Probe Icare TP01, 100 pcs/box

89 g

53 x 109 x 36 mm

540

Probe base

4 g

7 x 38 mm

559

Wrist strap with lock

4 g

10 x 10 x 270 mm

527

Aluminum case, Icare ic200

800 g

240 x 280 x 72 mm

7169

Battery cover & screw

6 g

110 x 25 x 12 mm

619

IOP pad, Icare ic200

38 g

50 x 53 x 16 mm

565

Silicone grip white

26 g

45 x 35 x 113 mm

566

Silicone grip green

26 g

45 x 35 x 113 mm

567

Silicone grip dark gray

26 g

45 x 35 x 113 mm

568

Silicone grip blue

26 g

45 x 35 x 113 mm

568A

Silicone grip pink

26 g

45 x 35 x 113 mm

548

Screwdriver

15 g

16 x 90 mm

577E

USB memory stick, Icare ic200

44 g

98 x 11 x 93 mm

544B

Probe base cover, Icare ic200

1 g

19 x 11 mm

Type: TA031 

Dimensions: 43 mm (W) x 104 mm (H) x 214 mm (L). 

Weight: 165 g (without batteries) 

Power supply: 4 x AA non-rechargeable batteries, 1.5 V alkaline LR6. 

Measurement range: 7-50 mmHg 

Accuracy: ±1.2 mm Hg (≤20 mmHg) and ±2.2 mm Hg (>20 mmHg). 

Repeatability (coefficient of variation): < 8 %. 

Accuracy of display: 0.1 mmHg. 

Display unit: Millimeters of mercury (mmHg). 

Mode of operation: continuous.

Operation environment:

Temperature: +10 °C to +35 °C

Relative humidity: 30 % to 90 %

Atmospheric pressure: 800 hPa to 1,060 hPa

Transport environment:

Temperature: -40 °C to +70 °C

Relative humidity: 10 % to 95 %

Atmospheric pressure: 500 hPa-1,060 hPa

Storage environment:

Temperature: -10 °C to +55 °C

Relative humidity: 10 % to 95 %

Atmospheric pressure: 700 hPa to 1,060 hPa

All manuals and user guides at all-guides.com

all-guides.com

Summary of Contents for ic200

Page 1: ...INSTRUCTION MANUAL ENGLISH All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 2: ...OBE 7 PROBE BASE INDICATOR LIGHT 7 MEASUREMENT 8 USER INTERFACE FUNCTIONS 11 BLUETOOTH 12 PRINTER 12 EXPORT 13 ERROR AND INFO MESSAGES 14 MEASUREMENT FLOW CHART 15 ACCESSORIES 16 TECHNICAL INFORMATION...

Page 3: ...nt in both eyes but in case either eye is inflamed or infected the healthy eye should be measured first Use probes taken only from the intact original packaging Re use of a probe could result in incor...

Page 4: ...f time it is recommended to remove the batteries from the battery compartment Removing the batteries will not affect the subsequent functioning of the tonometer Used probes cannot be recycled Dispose...

Page 5: ...ters covering the motion of the probe including deceleration and rebound time Using a proprietary algorithm the device is able to calculate the eye s IOP A displayed IOP reading is derived from the re...

Page 6: ...w the battery compartment locking screw with the screwdriver supplied Remove the battery compartment cover Insert a new set of four AA 1 5 V batteries LR6 Insert the batteries according to the figure...

Page 7: ...f the tonometer and probe device by emitting a red light if the device is too tilted to far up or a green light when the orientation of the device is acceptable Secondly the indicator light color chan...

Page 8: ...n STEP 2 You may choose to annotate the measurement result with the eye s side right left information Note that the device s default selection is NO eye side information Select between the OD right ey...

Page 9: ...17 8 17 4 18 2 mHg ME2 RECEI IF PRIN CONNE PRIN PRINTIN NOT FOUND PRINTER 12 PRINTER MENU If Printer is not found mmHg 12 6 OS 18 09 2017 07 37 45 AM 18 09 2017 07 37 45 AM 2 100 TO MENU 1 6 EYE OD O...

Page 10: ...successfully made from one eye you may measure the IOP in the other eye or make a repeat measurement in the same eye by repeating the steps 1 5 above The device does not automatically switch from one...

Page 11: ...2 6 3 6 4 6 5 6 0 LI LI LI LI 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU ME21 PAIR ME21 TEST ME21 UNPAIR BACK ME21 PAIRED TEST PRINTING REMOVED CONNECTION If PC Mobile Device is not found nne...

Page 12: ...11 1 12 1 13 1 15 12 0 12 2 12 3 12 4 12 6 12 4 12 6 12 8 12 8 12 5 12 1 IF PRINTER CONNECTED 1 14 1 16 2 2 2 3 1 17 1 18 1 8 1 7 3 2 3 3 17 6 23 9 20 8 16 5 17 6 17 4 23 8 1 4 1 5 EYE EYE OFF PATIEN...

Page 13: ...D 1 PATIENT ID ID02 PATIENT ID 5 1 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU If Printer is connected ME21 4 5 TO MENU BLUETOOTH Icare EXPORT SOUND SOUND SOUND SOUND SOUND SOUND OFF 12 2 12 3...

Page 14: ...4 14 6 14 3 14 5 14 7 14 9 14 8 14 10 14 11 14 ERRORs ALERTs 14 13 13 8 13 9 SERVICE 14 12 INFO 9 0 DATE 16 03 2017 TIME 17 15 10 0 11 0 10 1 10 2 10 3 8 0 8 1 8 3 8 4 LANGUAGE ENGLISH 8 2 1 2 3 4 BR...

Page 15: ...OAD LOAD mmHg mmHg mmHg mmHg mmHg mmHg 1 0 1 9 1 10 1 11 1 12 1 13 1 15 12 0 12 2 12 3 12 12 12 4 12 8 12 5 IF PRINTER CONNECTED 2 1 2 2 2 3 2 0 MEASURE 1 8 1 7 HISTORY mmHg 3 1 3 0 3 2 3 3 17 6 23 9...

Page 16: ...on installed The Icare Export will transfer the data into the Icare CLINIC software Operators can then access the data by using the Icare CLINIC software with a web browser in an internet connected de...

Page 17: ...s collar and probes are small objects and may be accidentally swallowed REPLACING THE PROBE BASE Change the probe base after six months If the error message Change is shown on the display screen more...

Page 18: ...ailure Non ionizing radiation Manufacturer Do not dispose of this product with household waste Send to appropriate facility for recovery and recycling EU WEEE European Union Directive for Waste of Ele...

Page 19: ...the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy...

Page 20: ...utions regarding EMC and needs to be installed and put into service according to EMC information provided in following tables Portable and mobile RF communications equipment can affect Icare ic200 TA0...

Page 21: ...terruption and voltage variations on power supply lines IEC 61000 4 11 0 UT for 0 5 cycle 1 phase 0 UT for 1 cycle 70 UT for 25 30 cycles 50 60 Hz 0 UT for 250 300 cycles 50 60 Hz NOT APPLICABLE NOT A...

Page 22: ...vicinity of equipment marked with the following symbol GUIDANCE AND MANUFACTURER S DECLARATION IEC 60601 1 2 2014 EDITION 4 0 ELECTROMAGNETIC IMMUNITY Icare ic200 TA031 is intended for use in a profes...

Page 23: ...ENGLISH Icare ic200 tonometer manual www icaretonometer com 23 NOTES All manuals and user guides at all guides com...

Reviews: