ENGLISH

Icare ic200 tonometer manual

www.icaretonometer.com

10

STEP 5

. Once six measurements have been successfully made, the tonometer will emit one long beep sound. The final IOP measurement is displayed in large 

digits, in mm Hg, inside a colored circle on the display screen. The colors indicate the quality of the IOP measurement. Green indicates “good” (a low variation 

of the observed parameters of probe motion during the four individual measurements used in the calculation of the final IOP), yellow indicates “acceptable” 

measurement quality.

If the variation in the measurements is too high, the tonometer will display the Repeat symbol on the display screen. A new measurement series can be initiated 

by pressing the Measure button once.

LIGHT

LIGHT

ME21

RECEIPT

OFF

TOO FAR

REPEAT

TOO NEAR

CHANGE

CHANGE

calibration

Sensor

CHANGE

CHANGE

14.0

14.1

14.2

14.4

14.6

14.3

14.5

14.7

14.9

14.8

14.10

14.11

REPEAT

LOAD

LOAD

mmHg

mmHg

mmHg

mmHg

mmHg

mmHg

14: ERRORs & ALERTs

14.13

1.0

1.9

1.10

1.11

1.12

1.13

1.15

12.0

12.2

13.0

13.1

13.2

13.5

13.1

13.6

13.7

13.8

13.9

13.3

13.3

13.4

12.3

12.4

12.6

12.7

12.8

12.9

12.10

12.4

12.6

12.8

12.8

12.5

12.5

12.1

IF PRINTER

CONNECTED

1.14

1.16

2.1

2.2

2.3

2.0

0.1

0.2

0.3

Document name: GUI Flowchart TA031

Document number: TA031-GUI

Revision: 04

Date: 11.14.2017

Author: Tomi Lonka (Desigence)

1.17

1.18

SERVICE

14.12

MEASURE

1.8

1.7

PATIENT ID

HISTORY

INFO

ID 1

mmHg

5.2

5.3

5.4

5.5

5.0

BLUETOOTH

740A

4.3

4.4

4.6

4.1

BLUETOOTH

OFF

4.2

BLUETOOTH

EXPORT

4.0

3.1

3.0

3.2

3.3

9.0

DATE

16.03.2017

TIME

17:15

10.0

11.0

10.1

10.2

10.3

8.0

8.1

8.3

8.4

LANGUAGE

ENGLISH

8.2

17

.6

23

.9

20

.8

16

.5

17

.6

17

.4

23

.8

1.1

1.2

1.4

1.5

1.3

EYE

EYE

OFF

PATIENT ID

BLUETOOTH

BLUETOOTH

RETURN

2.4

BACK

ID 1

PATIENT ID

ID02

PATIENT ID

5.1

6.2

6.3

6.4

6.5

6.0

LIGHT

6.1

LIGHT

LIGHT

LIGHT

7.1

7.2

7.3

7.4

BRIGHTNESS

7.0

BRIGHTNESS

BRIGHTNESS

BRIGHTNESS

LANGUAGE

KIELI

SPRÅK

SPRACHE

BRIGHTNESS

EN

FI

SV

DE

9.1

9.3

9.4

9.2

FORMAT

YEAR

MONTH

DAY

DMY

2017

03

16

11.1

FORMAT

HOURS

MINUTES

24 H

SW

17

15

2.5

RETURN

17.2 17.8

17.4

----

----

18.2

PRINT

PRINTING...

ID 9

mmHg

ME21

RECEIPT

IF PRINTER

CONNECTED

PRINT

PRINTING...

12: PRINTER MENU

13: PC/ MOBILE MENU

13: PC/ MOBILE MENU

NEW

PAIR

FOUND 1

SEARCHING...

NOT FOUND

PRINTER

ME21

PAIR

ME21

PAIR

ME21

TEST

ME21

UNPAIR

BACK

BACK

ME21

PAIRED

TEST

PRINTING...

REMOVED

CONNECTION

12: PRINTER MENU

If Printer is not found

If PC/ Mobile Device

is not found

If Printer is connected

If connection to PC/ Mobile device already exists

mmHg

17

.4

ID 9

1/100

OD

OS

OD

18/09/2017

07:37:45 AM

mmHg

12

.6

OS

18/09/2017

07:37:45 AM

18/09/2017

07:37:45 AM

2/100

TO MENU

ME21

4.5

TO MENU

ENGLISH

SOUMI

SVENSKA

DEUTSCH

1733RM001

1.00 A

SN

1733RM001

hh:mm

17

:mm

17:

15

DD.MM.YYYY

DD.MM.

2017

DD.

03

.2017

16

.03.2017

ERROR

PRINTER

ERROR

CONNECTION

RESTORED

CONNECTION

ME21

PAIRED

TEST

PRINTING...

PAIR

NEW

PAIR

740A

TEST

740A

UNPAIR

740A

WAITING...

DEVICE

WAITING...

DEVICE

PAIRED

740A

TBD

CONNECTION

CANCELLED

PAIRING

CANCELLED

012345

740A

PAIRED

740A

BLUETOOTH

Icare EXPORT

OD

OS

1.6

EYE

OD

OS

OD

OD

OD

OD

OD

OD

EYE

OD

OS

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

2/100

 3/100

SOUND

SOUND

SOUND

SOUND

SOUND

SOUND

16:36:10

16/03/2017

EYE

OD

OS

EYE

OS

OD

mmHg

16

.6

OD

mmHg

mmHg

17

.4

16

.5

mmHg

16

.5

ID 9

OD

LIGHT

LIGHT

ME21

RECEIPT

OFF

TOO FAR

REPEAT

TOO NEAR

CHANGE

CHANGE

calibration

Sensor

CHANGE

CHANGE

14.0

14.1

14.2

14.4

14.6

14.3

14.5

14.7

14.9

14.8

14.10

14.11

REPEAT

LOAD

LOAD

mmHg

mmHg

mmHg

mmHg

mmHg

mmHg

14: ERRORs & ALERTs

14.13

1.0

1.9

1.10

1.11

1.12

1.13

1.15

12.0

12.2

13.0

13.1

13.2

13.5

13.1

13.6

13.7

13.8

13.9

13.3

13.3

13.4

12.3

12.4

12.6

12.7

12.8

12.9

12.10

12.4

12.6

12.8

12.8

12.5

12.5

12.1

IF PRINTER

CONNECTED

1.14

1.16

2.1

2.2

2.3

2.0

0.1

0.2

0.3

Document name: GUI Flowchart TA031

Document number: TA031-GUI

Revision: 04

Date: 11.14.2017

Author: Tomi Lonka (Desigence)

1.17

1.18

SERVICE

14.12

MEASURE

1.8

1.7

PATIENT ID

HISTORY

INFO

ID 1

mmHg

5.2

5.3

5.4

5.5

5.0

BLUETOOTH

740A

4.3

4.4

4.6

4.1

BLUETOOTH

OFF

4.2

BLUETOOTH

EXPORT

4.0

3.1

3.0

3.2

3.3

9.0

DATE

16.03.2017

TIME

17:15

10.0

11.0

10.1

10.2

10.3

8.0

8.1

8.3

8.4

LANGUAGE

ENGLISH

8.2

17

.6

23

.9

20

.8

16

.5

17

.6

17

.4

23

.8

1.1

1.2

1.4

1.5

1.3

EYE

EYE

OFF

PATIENT ID

BLUETOOTH

BLUETOOTH

RETURN

2.4

BACK

ID 1

PATIENT ID

ID02

PATIENT ID

5.1

6.2

6.3

6.4

6.5

6.0

LIGHT

6.1

LIGHT

LIGHT

LIGHT

7.1

7.2

7.3

7.4

BRIGHTNESS

7.0

BRIGHTNESS

BRIGHTNESS

BRIGHTNESS

LANGUAGE

KIELI

SPRÅK

SPRACHE

BRIGHTNESS

EN

FI

SV

DE

9.1

9.3

9.4

9.2

FORMAT

YEAR

MONTH

DAY

DMY

2017

03

16

11.1

FORMAT

HOURS

MINUTES

24 H

SW

17

15

2.5

RETURN

17.2 17.8

17.4

----

----

18.2

PRINT

PRINTING...

ID 9

mmHg

ME21

RECEIPT

IF PRINTER

CONNECTED

PRINT

PRINTING...

12: PRINTER MENU

13: PC/ MOBILE MENU

13: PC/ MOBILE MENU

NEW

PAIR

FOUND 1

SEARCHING...

NOT FOUND

PRINTER

ME21

PAIR

ME21

PAIR

ME21

TEST

ME21

UNPAIR

BACK

BACK

ME21

PAIRED

TEST

PRINTING...

REMOVED

CONNECTION

12: PRINTER MENU

If Printer is not found

If PC/ Mobile Device

is not found

If Printer is connected

If connection to PC/ Mobile device already exists

mmHg

17

.4

ID 9

1/100

OD

OS

OD

18/09/2017

07:37:45 AM

mmHg

12

.6

OS

18/09/2017

07:37:45 AM

18/09/2017

07:37:45 AM

2/100

TO MENU

ME21

4.5

TO MENU

ENGLISH

SOUMI

SVENSKA

DEUTSCH

1733RM001

1.00 A

SN

1733RM001

hh:mm

17

:mm

17:

15

DD.MM.YYYY

DD.MM.

2017

DD.

03

.2017

16

.03.2017

ERROR

PRINTER

ERROR

CONNECTION

RESTORED

CONNECTION

ME21

PAIRED

TEST

PRINTING...

PAIR

NEW

PAIR

740A

TEST

740A

UNPAIR

740A

WAITING...

DEVICE

WAITING...

DEVICE

PAIRED

740A

TBD

CONNECTION

CANCELLED

PAIRING

CANCELLED

012345

740A

PAIRED

740A

BLUETOOTH

Icare EXPORT

OD

OS

1.6

EYE

OD

OS

OD

OD

OD

OD

OD

OD

EYE

OD

OS

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

2/100

 3/100

SOUND

SOUND

SOUND

SOUND

SOUND

SOUND

16:36:10

16/03/2017

EYE

OD

OS

EYE

OS

OD

mmHg

16

.6

OD

mmHg

mmHg

17

.4

16

.5

mmHg

16

.5

ID 9

OD

LIGHT

LIGHT

ME21

RECEIPT

OFF

TOO FAR

REPEAT

TOO NEAR

CHANGE

CHANGE

calibration

Sensor

CHANGE

CHANGE

14.0

14.1

14.2

14.4

14.6

14.3

14.5

14.7

14.9

14.8

14.10

14.11

REPEAT

LOAD

LOAD

mmHg

mmHg

mmHg

mmHg

mmHg

mmHg

14: ERRORs & ALERTs

14.13

1.0

1.9

1.10

1.11

1.12

1.13

1.15

12.0

12.2

13.0

13.1

13.2

13.5

13.1

13.6

13.7

13.8

13.9

13.3

13.3

13.4

12.3

12.4

12.6

12.7

12.8

12.9

12.10

12.4

12.6

12.8

12.8

12.5

12.5

12.1

IF PRINTER

CONNECTED

1.14

1.16

2.1

2.2

2.3

2.0

0.1

0.2

0.3

Document name: GUI Flowchart TA031

Document number: TA031-GUI

Revision: 04

Date: 11.14.2017

Author: Tomi Lonka (Desigence)

1.17

1.18

SERVICE

14.12

MEASURE

1.8

1.7

PATIENT ID

HISTORY

INFO

ID 1

mmHg

5.2

5.3

5.4

5.5

5.0

BLUETOOTH

740A

4.3

4.4

4.6

4.1

BLUETOOTH

OFF

4.2

BLUETOOTH

EXPORT

4.0

3.1

3.0

3.2

3.3

9.0

DATE

16.03.2017

TIME

17:15

10.0

11.0

10.1

10.2

10.3

8.0

8.1

8.3

8.4

LANGUAGE

ENGLISH

8.2

17

.6

23

.9

20

.8

16

.5

17

.6

17

.4

23

.8

1.1

1.2

1.4

1.5

1.3

EYE

EYE

OFF

PATIENT ID

BLUETOOTH

BLUETOOTH

RETURN

2.4

BACK

ID 1

PATIENT ID

ID02

PATIENT ID

5.1

6.2

6.3

6.4

6.5

6.0

LIGHT

6.1

LIGHT

LIGHT

LIGHT

7.1

7.2

7.3

7.4

BRIGHTNESS

7.0

BRIGHTNESS

BRIGHTNESS

BRIGHTNESS

LANGUAGE

KIELI

SPRÅK

SPRACHE

BRIGHTNESS

EN

FI

SV

DE

9.1

9.3

9.4

9.2

FORMAT

YEAR

MONTH

DAY

DMY

2017

03

16

11.1

FORMAT

HOURS

MINUTES

24 H

SW

17

15

2.5

RETURN

17.2 17.8

17.4

----

----

18.2

PRINT

PRINTING...

ID 9

mmHg

ME21

RECEIPT

IF PRINTER

CONNECTED

PRINT

PRINTING...

12: PRINTER MENU

13: PC/ MOBILE MENU

13: PC/ MOBILE MENU

NEW

PAIR

FOUND 1

SEARCHING...

NOT FOUND

PRINTER

ME21

PAIR

ME21

PAIR

ME21

TEST

ME21

UNPAIR

BACK

BACK

ME21

PAIRED

TEST

PRINTING...

REMOVED

CONNECTION

12: PRINTER MENU

If Printer is not found

If PC/ Mobile Device

is not found

If Printer is connected

If connection to PC/ Mobile device already exists

mmHg

17

.4

ID 9

1/100

OD

OS

OD

18/09/2017

07:37:45 AM

mmHg

12

.6

OS

18/09/2017

07:37:45 AM

18/09/2017

07:37:45 AM

2/100

TO MENU

ME21

4.5

TO MENU

ENGLISH

SOUMI

SVENSKA

DEUTSCH

1733RM001

1.00 A

SN

1733RM001

hh:mm

17

:mm

17:

15

DD.MM.YYYY

DD.MM.

2017

DD.

03

.2017

16

.03.2017

ERROR

PRINTER

ERROR

CONNECTION

RESTORED

CONNECTION

ME21

PAIRED

TEST

PRINTING...

PAIR

NEW

PAIR

740A

TEST

740A

UNPAIR

740A

WAITING...

DEVICE

WAITING...

DEVICE

PAIRED

740A

TBD

CONNECTION

CANCELLED

PAIRING

CANCELLED

012345

740A

PAIRED

740A

BLUETOOTH

Icare EXPORT

OD

OS

1.6

EYE

OD

OS

OD

OD

OD

OD

OD

OD

EYE

OD

OS

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

ID 9

2/100

 3/100

SOUND

SOUND

SOUND

SOUND

SOUND

SOUND

16:36:10

16/03/2017

EYE

OD

OS

EYE

OS

OD

mmHg

16

.6

OD

mmHg

mmHg

17

.4

16

.5

mmHg

16

.5

ID 9

OD

The values from the first to the fifth value displayed before the sixth value are running average values. The sixth value is the final IOP value which is calculated 

from the best four individual measurements (the worst two individual measurements are discarded).

STEP 6

. After an IOP measurement has been successfully made from one eye, you may measure the IOP in the other eye (or make a repeat measurement in the 

same eye) by repeating the steps 1-5 above. The device does not automatically switch from one eye to another after one eye is measured, e.g. from OD to OS.

When you have finished your IOP measurement session, hold the device so that the probe is horizontally or slightly downward tilted, press the Select button for 

three seconds to turn the tonometer off. The probe comes out from the probe base and you can remove it. Discard the probe (according to instructions). Retrieve 

the probe base cover and place over the probe base.

Note: When the tonometer is not in use, always keep the probe base covered to protect the probe base from contamination.

If you doubt the validity of any of the tonometer’s displayed measurements of IOP (for example, if you suspect that the probe missed the central cornea or made 

contact with the eyelid), it is recommended that you repeat the measurement. In addition, if you observe an unusually high or low displayed value of IOP, it is 

recommended that you make another measurement, either with the Icare tonometer or using an alternative method in order to verify the unusual reading.

If you are unable to complete six successful measurements in a sequence, the measurement process can be terminated by pressing the Select button once. In 

such an instance, the results of the measurement attempt can be viewed in the device’s HISTORY menu. Note that in cases of unfinished measurements, IOP data 

from the individual steps are displayed with no indication of measurement validity.

All manuals and user guides at all-guides.com

Summary of Contents for ic200

Page 1: ...INSTRUCTION MANUAL ENGLISH All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 2: ...OBE 7 PROBE BASE INDICATOR LIGHT 7 MEASUREMENT 8 USER INTERFACE FUNCTIONS 11 BLUETOOTH 12 PRINTER 12 EXPORT 13 ERROR AND INFO MESSAGES 14 MEASUREMENT FLOW CHART 15 ACCESSORIES 16 TECHNICAL INFORMATION...

Page 3: ...nt in both eyes but in case either eye is inflamed or infected the healthy eye should be measured first Use probes taken only from the intact original packaging Re use of a probe could result in incor...

Page 4: ...f time it is recommended to remove the batteries from the battery compartment Removing the batteries will not affect the subsequent functioning of the tonometer Used probes cannot be recycled Dispose...

Page 5: ...ters covering the motion of the probe including deceleration and rebound time Using a proprietary algorithm the device is able to calculate the eye s IOP A displayed IOP reading is derived from the re...

Page 6: ...w the battery compartment locking screw with the screwdriver supplied Remove the battery compartment cover Insert a new set of four AA 1 5 V batteries LR6 Insert the batteries according to the figure...

Page 7: ...f the tonometer and probe device by emitting a red light if the device is too tilted to far up or a green light when the orientation of the device is acceptable Secondly the indicator light color chan...

Page 8: ...n STEP 2 You may choose to annotate the measurement result with the eye s side right left information Note that the device s default selection is NO eye side information Select between the OD right ey...

Page 9: ...17 8 17 4 18 2 mHg ME2 RECEI IF PRIN CONNE PRIN PRINTIN NOT FOUND PRINTER 12 PRINTER MENU If Printer is not found mmHg 12 6 OS 18 09 2017 07 37 45 AM 18 09 2017 07 37 45 AM 2 100 TO MENU 1 6 EYE OD O...

Page 10: ...successfully made from one eye you may measure the IOP in the other eye or make a repeat measurement in the same eye by repeating the steps 1 5 above The device does not automatically switch from one...

Page 11: ...2 6 3 6 4 6 5 6 0 LI LI LI LI 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU ME21 PAIR ME21 TEST ME21 UNPAIR BACK ME21 PAIRED TEST PRINTING REMOVED CONNECTION If PC Mobile Device is not found nne...

Page 12: ...11 1 12 1 13 1 15 12 0 12 2 12 3 12 4 12 6 12 4 12 6 12 8 12 8 12 5 12 1 IF PRINTER CONNECTED 1 14 1 16 2 2 2 3 1 17 1 18 1 8 1 7 3 2 3 3 17 6 23 9 20 8 16 5 17 6 17 4 23 8 1 4 1 5 EYE EYE OFF PATIEN...

Page 13: ...D 1 PATIENT ID ID02 PATIENT ID 5 1 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU If Printer is connected ME21 4 5 TO MENU BLUETOOTH Icare EXPORT SOUND SOUND SOUND SOUND SOUND SOUND OFF 12 2 12 3...

Page 14: ...4 14 6 14 3 14 5 14 7 14 9 14 8 14 10 14 11 14 ERRORs ALERTs 14 13 13 8 13 9 SERVICE 14 12 INFO 9 0 DATE 16 03 2017 TIME 17 15 10 0 11 0 10 1 10 2 10 3 8 0 8 1 8 3 8 4 LANGUAGE ENGLISH 8 2 1 2 3 4 BR...

Page 15: ...OAD LOAD mmHg mmHg mmHg mmHg mmHg mmHg 1 0 1 9 1 10 1 11 1 12 1 13 1 15 12 0 12 2 12 3 12 12 12 4 12 8 12 5 IF PRINTER CONNECTED 2 1 2 2 2 3 2 0 MEASURE 1 8 1 7 HISTORY mmHg 3 1 3 0 3 2 3 3 17 6 23 9...

Page 16: ...on installed The Icare Export will transfer the data into the Icare CLINIC software Operators can then access the data by using the Icare CLINIC software with a web browser in an internet connected de...

Page 17: ...s collar and probes are small objects and may be accidentally swallowed REPLACING THE PROBE BASE Change the probe base after six months If the error message Change is shown on the display screen more...

Page 18: ...ailure Non ionizing radiation Manufacturer Do not dispose of this product with household waste Send to appropriate facility for recovery and recycling EU WEEE European Union Directive for Waste of Ele...

Page 19: ...the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy...

Page 20: ...utions regarding EMC and needs to be installed and put into service according to EMC information provided in following tables Portable and mobile RF communications equipment can affect Icare ic200 TA0...