GUIDANCE AND MANUFACTURER’S DECLARATION IEC 60601-1-2:2014; EDITION 4.0 - ELECTROMAGNETIC IMMUNITY
Icare ic200 (TA031) is intended for use in a professional healthcare environment with electromagnetic characteristics specified below.
The user of the Icare ic200 Tonometer (TA031) should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Conducted
disturbances
induced by RF
fields IEC 61000-
4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz 80 % AM at 1 kHz
NOT APPLICABLE
NOT APPLICABLE
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 V/m
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Icare ic200 (TA031) including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Interference may occur in the vicinity of equipment marked with the following
symbol:
GUIDANCE AND MANUFACTURER’S DECLARATION IEC 60601-1-2: 2014; EDITION 4.0 - ELECTROMAGNETIC IMMUNITY
Icare ic200 (TA031) is intended for use in a professional healthcare environment with electromagnetic characteristics specified below.
The user of the Icare ic200 Tonometer (TA031) should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz 27 V/m;
PM 50%; 18 Hz
430 - 470 MHz 28 V/m; (FM
±5 kHz, 1 kHz sine) PM; 18 Hz
704 - 787 MHz 9 V/m;
PM 50%; 217 Hz
800 - 960 MHz 28 V/m;
PM 50%; 18 Hz
1700 - 1990 MHz 28 V/m;
PM 50%; 217 Hz
2400 - 2570 MHz 28 V/m;
PM 50%; 217 Hz
5100 - 5800 MHz 9 V/m;
PM 50%; 217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Icare ic200 (TA031) including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Interference may occur in the vicinity of equipment marked with the following
symbol: