ENGLISH

Icare ic200 tonometer manual  

www.icaretonometer.com

5

INTENDED USE

The Icare ic200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.

The Icare ic200 tonometer is intended to be used by healthcare professionals in a professional healthcare environment. No special skills or

training are needed to use the device.

INTRODUCTION

The Icare ic200 tonometer is based on a patented, induction-based rebound method, which allows intraocular pressure (IOP) to be measured accurately, rapidly 

and without an anesthetic.

The Icare ic200 tonometer allows for a patient’s IOP to be measured for patients in the supine position as well as in upright (sitting or standing) positions.

With the Icare rebound measurement method, a miniature, light-weight probe is launched in a direction perpendicular to the surface of the center of the cornea 

of the eye. The probe is composed of a medical grade plastic tip and a metal shaft. The metal shaft is magnetized prior to measurement. During a measurement, 

the probe can be considered to act like a moving magnet that induces an electric signal in the surrounding coil allowing for highly accurate measurement of the 

probe’s motion. After launching, the probe briefly makes contact with the cornea and bounces back. The tonometer records multiple parameters covering the 

motion of the probe, including deceleration and rebound time. Using a proprietary algorithm, the device is able to calculate the eye’s IOP.

A displayed IOP reading is derived from the results from a sequence of six individual IOP measurements and calculations, made each of the six times the probe 

hits the cornea and rebounds. The displayed IOP measurement is also stored in the tonometer’s memory for later retrieval.

The Icare ic200 tonometer incorporates a Bluetooth® module which allows wireless connectivity to Bluetooth-supported printers or for data transfer.

No part of the tonometer or probes contain natural rubber latex.

PACKAGE CONTENTS

 PRECAUTION

When removing the device from its packaging, visually inspect the tonometer for any external damage, particularly for possible damage to the device casing. 

If you suspect damage to the tonometer, contact the manufacturer or distributor of the device.

The Icare ic200 package contains:

•  Icare ic200 tonometer 
•  4 x AA 1.5 V batteries
•  Box of 100 probes
•  wrist strap
•  silicone grip

•  IOP pad for writing down results
•  aluminum case
•  screwdriver
•  probe base cover
•  spare probe base

• printed quick guide
• USB drive including instruction manuals
• printed instruction manuals (EU only)
• warranty card

FEATURES AND PARTS OF THE TONOMETER

1.

Forehead support 

2.

Probe base

3.

Locking collar

4.

Display screen

5.

Adjustment wheel for forehead 

support

6.

User interface Navigation buttons

7.

Select 

button

8.

Measure

 button

1.

2.

3.

4.

6.

7.
8.

5.

All manuals and user guides at all-guides.com

Summary of Contents for ic200

Page 1: ...INSTRUCTION MANUAL ENGLISH All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 2: ...OBE 7 PROBE BASE INDICATOR LIGHT 7 MEASUREMENT 8 USER INTERFACE FUNCTIONS 11 BLUETOOTH 12 PRINTER 12 EXPORT 13 ERROR AND INFO MESSAGES 14 MEASUREMENT FLOW CHART 15 ACCESSORIES 16 TECHNICAL INFORMATION...

Page 3: ...nt in both eyes but in case either eye is inflamed or infected the healthy eye should be measured first Use probes taken only from the intact original packaging Re use of a probe could result in incor...

Page 4: ...f time it is recommended to remove the batteries from the battery compartment Removing the batteries will not affect the subsequent functioning of the tonometer Used probes cannot be recycled Dispose...

Page 5: ...ters covering the motion of the probe including deceleration and rebound time Using a proprietary algorithm the device is able to calculate the eye s IOP A displayed IOP reading is derived from the re...

Page 6: ...w the battery compartment locking screw with the screwdriver supplied Remove the battery compartment cover Insert a new set of four AA 1 5 V batteries LR6 Insert the batteries according to the figure...

Page 7: ...f the tonometer and probe device by emitting a red light if the device is too tilted to far up or a green light when the orientation of the device is acceptable Secondly the indicator light color chan...

Page 8: ...n STEP 2 You may choose to annotate the measurement result with the eye s side right left information Note that the device s default selection is NO eye side information Select between the OD right ey...

Page 9: ...17 8 17 4 18 2 mHg ME2 RECEI IF PRIN CONNE PRIN PRINTIN NOT FOUND PRINTER 12 PRINTER MENU If Printer is not found mmHg 12 6 OS 18 09 2017 07 37 45 AM 18 09 2017 07 37 45 AM 2 100 TO MENU 1 6 EYE OD O...

Page 10: ...successfully made from one eye you may measure the IOP in the other eye or make a repeat measurement in the same eye by repeating the steps 1 5 above The device does not automatically switch from one...

Page 11: ...2 6 3 6 4 6 5 6 0 LI LI LI LI 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU ME21 PAIR ME21 TEST ME21 UNPAIR BACK ME21 PAIRED TEST PRINTING REMOVED CONNECTION If PC Mobile Device is not found nne...

Page 12: ...11 1 12 1 13 1 15 12 0 12 2 12 3 12 4 12 6 12 4 12 6 12 8 12 8 12 5 12 1 IF PRINTER CONNECTED 1 14 1 16 2 2 2 3 1 17 1 18 1 8 1 7 3 2 3 3 17 6 23 9 20 8 16 5 17 6 17 4 23 8 1 4 1 5 EYE EYE OFF PATIEN...

Page 13: ...D 1 PATIENT ID ID02 PATIENT ID 5 1 12 PRINTER MENU 13 PC MOBILE MENU 13 PC MOBILE MENU If Printer is connected ME21 4 5 TO MENU BLUETOOTH Icare EXPORT SOUND SOUND SOUND SOUND SOUND SOUND OFF 12 2 12 3...

Page 14: ...4 14 6 14 3 14 5 14 7 14 9 14 8 14 10 14 11 14 ERRORs ALERTs 14 13 13 8 13 9 SERVICE 14 12 INFO 9 0 DATE 16 03 2017 TIME 17 15 10 0 11 0 10 1 10 2 10 3 8 0 8 1 8 3 8 4 LANGUAGE ENGLISH 8 2 1 2 3 4 BR...

Page 15: ...OAD LOAD mmHg mmHg mmHg mmHg mmHg mmHg 1 0 1 9 1 10 1 11 1 12 1 13 1 15 12 0 12 2 12 3 12 12 12 4 12 8 12 5 IF PRINTER CONNECTED 2 1 2 2 2 3 2 0 MEASURE 1 8 1 7 HISTORY mmHg 3 1 3 0 3 2 3 3 17 6 23 9...

Page 16: ...on installed The Icare Export will transfer the data into the Icare CLINIC software Operators can then access the data by using the Icare CLINIC software with a web browser in an internet connected de...

Page 17: ...s collar and probes are small objects and may be accidentally swallowed REPLACING THE PROBE BASE Change the probe base after six months If the error message Change is shown on the display screen more...

Page 18: ...ailure Non ionizing radiation Manufacturer Do not dispose of this product with household waste Send to appropriate facility for recovery and recycling EU WEEE European Union Directive for Waste of Ele...

Page 19: ...the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy...

Page 20: ...utions regarding EMC and needs to be installed and put into service according to EMC information provided in following tables Portable and mobile RF communications equipment can affect Icare ic200 TA0...

Page 21: ...terruption and voltage variations on power supply lines IEC 61000 4 11 0 UT for 0 5 cycle 1 phase 0 UT for 1 cycle 70 UT for 25 30 cycles 50 60 Hz 0 UT for 250 300 cycles 50 60 Hz NOT APPLICABLE NOT A...

Page 22: ...vicinity of equipment marked with the following symbol GUIDANCE AND MANUFACTURER S DECLARATION IEC 60601 1 2 2014 EDITION 4 0 ELECTROMAGNETIC IMMUNITY Icare ic200 TA031 is intended for use in a profes...

Page 23: ...ENGLISH Icare ic200 tonometer manual www icaretonometer com 23 NOTES All manuals and user guides at all guides com...

Reviews: