In this example, the two measurements for meter no. 7 differ more then 0,1. A third water
test is performed. All three differences are within the 0,2 limit, so no futher action needed.
o
Based on de water test differences a new
G
parameter can be calculated (
). If the new
G
parameter differs more then 5 (meaning >2%
deviation and adjustment), two additional water tests for that specific meter are
required after the adjustment of G in order to set references (in this example meter
4).
o
Set de new
G
parameter in the individual meter using the
G
adjustment procedure
(see links below)
For meter no. 4 the adjustment is more than 2% and two new water tests are needed.
In this example the new water test looks like:
To adjust the
G
parameter valua follow the procedure in paragraph
or via the AfiControl interface mentioned in paragraph
“G” and/or “H” parameter value via communication
For the periodic routine test, the following results, following this example, have to be
handed over to the persons/organizations involved in routine testing (reference values for
the next annual routine test :
Form with the calculated reference values for the annual routine test based on the
calibration test.
Conduc-
Conduc-
Conduc-
Calculated
Meter G-value
Display
Unster
Diff.
tivity
afwijking
Display Unster
Diff.
tivity
Display
Unster
Diff. tivity
Diff
Deviation%
G- value
1
53
12,20
11,44
0,76
11,7
12,32 11,48
0,84
11,8
0,80
-0,2%
53
2
53
12,30
11,56
0,74
11,6
12,26 11,46
0,80
11,8
0,77
-0,5%
54
3
53
12,25
11,55
0,70
11,6
12,20 11,48
0,72
11,7
0,71
-0,9%
55
4
53
12,28
11,20
1,08
11,6
12,32 11,20
1,12
11,6
1,10
2,2%
47
5
53
12,30
11,48
0,82
11,5
12,40 11,60
0,80
11,6
0,81
-0,1%
53
6
53
12,44
11,68
0,76
11,8
12,30 11,48
0,82
11,9
0,79
-0,3%
53
7
53
12,18
11,26
0,92
11,7
12,46 11,66
0,80
11,8
12,40 11,52 0,88
11,7
0,87
0,3%
53
8
53
12,20
11,40
0,80
11,7
12,28 11,46
0,82
11,7
0,81
-0,1%
53
9
53
12,24
11,46
0,78
11,6
12,26 11,50
0,76
11,6
0,77
-0,5%
54
10
53
12,16
11,36
0,80
11,5
12,20 11,34
0,86
11,6
0,83
0,0%
53
0,82
Farm reference wil be:
0,83
Avarage
Avarage water test diff.
Pre water test
2e water test
3e water test
Vacuum level (kPa)
42,0
Parlor: Harringbone (2x5)
Date: 30-7-2018
Test person Joe
three within 0,2 ( else replace faulty measurements)
two measurements with mutual difference max. 0,1 kg
Afimilk MPC Setting reference values by water test
Farm:
1265897
Technical check
Farm tesing
Hygienic check
Cow road
Performer:
Cow city
Conduc-
Conduc- Meter
Meter G-value
Display
Unster
Diff.
tivity
afwijking
Display Unster
Diff.
tivity
Reference
4
47
12,10
11,40
0,70
11,8
12,16 11,48
0,68
11,8
0,69
1e water test
2e water test