12
AfiMilk MPC
Installation Manual
To adjust the
G
parameter valua follow the procedure in paragraph
or via the AfiControl interface mentioned in paragraph
and/or H parameter value via communication
7.3.1.1
Equipment and water test preparation and description
7.3.1.1.1
Equipment required
For performing the AfiMilk MPC water test, the following
equipment is used:
Milking bucket
equipped with 7/8” inlet and
vacuum ports
Manual vacuum shut off valve
placed between the milk line
and the hose connected to the collection bucket.
Suction pipe
with an inlet hole of 3.5mm, which allows a 4 ± 1 LPM (Liter per Minute)
flow of water. The entire set should be
tested and verified for the defined flow
rate (see testing the equipment
below). This part will replace the
milking claw for the purpose of the
test.
Make sure you work accurate with collecting all the data you need.
This includes i.e. missing data due to missing cow identification, milk thru
the meter but not in the bulk tank and milk substraction out of bulk tank.
.
IMPORTANT