21
an error is displayed 11
Error
Fault/Cause
Solution/Explanation
The pulse frequency
is higher or lower
than the average
value.
The unit switches off
automatically.
The unit continues
to pump up while
the reading is being
taken.
You have moved your arm or body
briefly during the reading or you
have spoken.
You have taken the reading
immediately after physical exertion.
This is built into the system.
This is normal. If the user’s blood
pressure that is being measured is
higher than the initial value of the
sleeve pressure, the unit will
automatically continue to pump.
The pressure build-up will occur in
stages of 20 mmHg, until a suitable
pressure for the reading has been
reached.
The sleeve is not correctly fitted or is
not tight enough.
Do not move while the reading is
being taken. Do not talk.
Do not take any reading immediately
after physical exertion.
Press the
START/STOP
key again
and re-take the reading.
Fit the sleeve correctly
(see Ch. 9.1, pp. 13-14).
BP 603 GB_neu 04.01.2005 10:45 Uhr Seite 21
Summary of Contents for BP 603
Page 2: ...BP 603 D_neu 04 01 2005 10 54 Uhr Seite 2 ...
Page 28: ...www ibp info de BP 603 D_neu 04 01 2005 10 55 Uhr Seite 26 ...
Page 30: ...BP 603 GB_neu 04 01 2005 10 45 Uhr Seite 2 ...
Page 56: ...www ibp info de BP 603 GB_neu 04 01 2005 10 45 Uhr Seite 26 ...
Page 58: ...BP 603 F_neu 04 01 2005 10 48 Uhr Seite 2 ...
Page 84: ...www ibp info de BP 603 F_neu 04 01 2005 10 48 Uhr Seite 26 ...
Page 86: ...BP 603 NL_neu 04 01 2005 10 44 Uhr Seite 2 ...
Page 112: ...www ibp info de BP 603 NL_neu 04 01 2005 10 44 Uhr Seite 26 ...
Page 114: ...BP 603 E_neu 04 01 2005 10 51 Uhr Seite 2 ...
Page 140: ...www ibp info de BP 603 E_neu 04 01 2005 10 51 Uhr Seite 26 ...