Manual ibaPADU-8-ICP
8 Issue
1.7
3
Scope of delivery
After unpacking check the completeness and intactness of the delivery.
The scope of delivery includes:
ibaPADU-8-ICP Device with Phoenix terminal blocks
ibaPADU-8-ICP Documentation
2 each Phoenix terminal blocks for the ICP
®
channels
1 each Phoenix terminal block for the binary channels
4
Safety instructions
4.1
Designated use
The device is electrical equipment. It may be used only in the following applications:
Measurement data logging and analysis
Applications of ibaSoftware products (ibaLogic, ibaPDA-V6 etc.)
The device may not be operated in mains supply circuits!
4.2
Special advices
Important note
Do not open the device!
There are no serviceable parts inside the device.
Opening the device will void the warranty.
Note
Sensor cable shield should be connected only on the ibaPADU-8-ICP side in order to
avoid ground loops with the sensor housing.
Note
Always ground the DIN rail on which the device is installed.
Note
Cleaning
To clean the device, use a dry or slightly moistened cloth. A note specifies special re-
quirements or actions to be observed.