Important Information: Read This First!

To receive safe and optimum system benefits, please read the 

entire manual contents before using the system. Please note 

the following instructions:

•   Do not use the system for diagnosis or screening of diabetes.

•  Use only fresh capillary whole blood samples for testing.

•   Alternative site and fingertip test results may differ 

significantly due to rapid changes in the blood glucose level 

after eating, insulin injection or exercise.

Intended Use

The CareSens N Voice Blood Glucose Monitoring System is 

intended for the quantitative measurement of glucose in fresh 

capillary whole blood samples drawn from the fingertips and 

alternative sites such as the forearm, palm, thigh, and calf. 

Alternative site testing should be used only during steady-state 

blood glucose conditions. The CareSens N Voice Blood Glucose 

Monitoring System is intended for self testing outside the body 

(

in vitro

) by people with diabetes at home as an aid to monitor 

the effectiveness of diabetes control. The system is intended to 

be used by a single person and should not be shared. It is not 

intended for use on neonates and is not for the diagnosis or 

screening of diabetes.

The CareSens N Single Blood Glucose Test Strips are for use with 

the CareSens N Voice Blood Glucose Meter to quantitatively 

measure glucose in fresh capillary whole blood samples drawn 

from the fingertips and alternative sites.

The CareSens Control Solutions are for use with the CareSens N 

Voice Meter and CareSens N Single Test Strips to check that the 

meter and the test strips are working together properly and that 

the test is performing correctly.

•   The CareSens N Voice Blood Glucose Monitoring System is 

intended for 

in vitro

 (outside the body) diagnostic use only 

and for the quantitative measurement of glucose in capillary 

whole blood.

•   The CareSens N Voice Blood Glucose Monitoring System is 

intended for use by a single user only. Do not share the meter 

and lancing device with another person.

•   The CareSens N Voice Blood Glucose Meter is designed to 

minimize code related errors in monitoring by using the 

no-coding function.

•   The glucose in the blood sample mixes with special chemicals 

on the test strip to produce a small electrical current. The 

CareSens N Voice Meter converts this electrical current to 

glucose concentration.

•   The CareSens N Voice Blood Glucose Meter should be used 

only with the CareSens N Single Test Strip.

•   An abnormally high or low red blood cell count (hematocrit 

level over 60% or below 20%) may produce inaccurate results.

•   Inaccurate results may occur in severely hypotensive 

(having low blood pressure) individuals or patients in shock. 

Inaccurate low results may occur for individuals experiencing a 

hyperglycemic (high blood sugar) or hyperosmolar state, with 

or without ketosis. Critically ill patients should not be tested 

with blood glucose meters.

•   All parts of the kit are considered biohazardous and can 

potentially transmit infectious diseases, even after you have 

performed pre-cleaning and disinfection.

•    For questions or concerns, contact Customer Service: 1-800-

429-5001 (9 am - 9 pm EST, Mon - Sat). At all other times 

or in case of emergency, please contact your healthcare 

professional or emergency medical response.

Important Health-Related Information

Summary of Contents for CareSens N Voice

Page 1: ...r s lo Sistema de Monitoreo de Glucosa en Sangre Manual del Usuario Funci n de voz L No Codificaci n Segundos For single user home use only Blood Glucose Monitoring System User s Manual No Coding Talk...

Page 2: ...professional or emergency medical response Bien venido al Sistema de Monitoreo de Glucosa CareSens N Voice Gracias por elegir el Sistema de Monitoreo de Glucosa en Sangre CareSens N Voice El sistema l...

Page 3: ...d Glucose Testing Using the Lancing Device 23 Preparing the Lancing Device 24 Preparing the Meter and Test Strip 25 Flagging Post meal Test Results 26 Applying the Blood Sample 27 Discarding Used Lanc...

Page 4: ...rips to check that the meter and the test strips are working together properly and that the test is performing correctly The CareSens N Voice Blood Glucose Monitoring System is intended for in vitro o...

Page 5: ...N Voice Blood Glucose Meter User s Manual Quick Reference Guide Batteries 2 CareSens N Voice System may include the following items CareSens N Single Blood Glucose Test Strips 10 Lancing Device Lancet...

Page 6: ...ever you may need to reset your meter settings See page 13 CareSens N Single Blood Glucose Test Strip The CareSens N Voice Blood Glucose Monitoring System measures blood glucose quickly and accurately...

Page 7: ...er test strips in any way For detailed storage and usage information refer to the CareSens N Single Test Strip package insert Keep the meter and testing supplies away from young children Note You can...

Page 8: ...date the meter settings such as time and date before using your meter or when you change the meter batteries Press and hold the S button for 3 seconds to enter the SET mode While you are setting the m...

Page 9: ...ion is on Press or button until blinking number shows the correct date Press the S button to confirm the date and set the time Step 6 Setting the Time Format The meter can be set in the AM PM 12 hour...

Page 10: ...20 mg dL Step 10 Turning on the Hypoglycemia Indicator When you enter the hypoglycemia indicator setting mode the symbol and HYPo will blink If the talking function is on the meter will say Set hypogl...

Page 11: ...using the control solution shake the bottle discard the first 1 or 2 drops and wipe the tip clean Close the control solution bottle tightly and store at a temperature of 46 86 F 8 30 C You should do...

Page 12: ...solution test result by pressing and holding the button for 3 seconds until the ctl appears on the display When the ctl is displayed the result is stored in the meter s memory but is not included in t...

Page 13: ...ned plus three 3 months When the control solution is contaminated Situations Actions Repeat the control solution test by referring to the Notes on page 18 Discard the used control solution and repeat...

Page 14: ...h of penetration by rotating the top portion of the adjustable tip until the setting number matches the arrow Step 5 Cock the lancing device by holding the body in one hand and pulling on the sleeve w...

Page 15: ...ng the blood sample or when the test result is displayed just after testing Once the test result is saved with the post meal flag it cannot be deleted Step 9 After the symbol appears on the screen app...

Page 16: ...of the symbol appearing on the screen If the meter turns off press S button or button and apply the blood sample after the symbol appears on the screen Step 1 Unscrew the lancing device tip Step 2 Pla...

Page 17: ...device changes color Then press the release button while continuing to apply pressure Keep holding the lancing device against your skin until sufficient at least 0 5 L blood is drawn Carefully lift t...

Page 18: ...e to contact with hair or with lines on your palm do not use that sample Try puncturing again in a smoother area Note If messages for hyperglycemia or hypoglycemia are displayed even if you do not hav...

Page 19: ...isplayed when the data cable connects the meter with computer For more information contact Customer Service at 1 800 429 5001 Meter Memory Step 1 Press the S or button to turn on the meter If the talk...

Page 20: ...previously To turn off the meter press the S button button button Post meal average 1 7 14 30 90 days button button Pre meal average 1 7 14 30 90 days Step 2 Use the button to scroll through the store...

Page 21: ...ut symbol being displayed Setting the Alarm Function Setting the Post meal Alarm PP2 alarm Four types of alarms can be set in the CareSens N Voice Meter one post meal alarm PP2 alarm and three time se...

Page 22: ...d lancing device should be disinfected prior to use by the second person Pre cleaning procedure is needed to clean dirt as well as blood and other body fluids on the exterior of the meter and lancing...

Page 23: ...the disinfectant product please contact the manufacturer or visit their website as listed above Name Manufacturer EPAregistration number Active ingredients Clorox Germicidal Wipes Clorox Professional...

Page 24: ...nfectants because other chemicals have not been validated and may damage the meter Do Not get fluids inside the meter through the test strip port data transmission port or battery compartment Never im...

Page 25: ...n or technical assistance by calling our Customer Service Center 1 800 429 5001 9 am 9 pm EST Monday Saturday Message What is said Error 1 Used test strip Error 2 The sample has been applied before th...

Page 26: ...Customer Service 1 800 429 5001 9 am 9 pm EST Monday Saturday Message What is said What it means What to do Error 6 Error 7 Error 8 Error 9 Meter failure There is a problem with the meter Do not use t...

Page 27: ...pm EST Monday Saturday Within 5 Within 10 Within 15 Within 20 54 93 58 82 93 88 93 93 100 93 93 100 Within 5mg dL 14 17 82 17 17 100 17 17 100 Within 10mg dL Within 15mg dL Performance Characteristics...

Page 28: ...omparison of results using various AST sites with the results of YSI measurements Alternative site test results for glucose concentration 75 mg dL Within 5mg dL Within 10mg dL Within 15mg dL Fingertip...

Page 29: ...r handling tampering use or service of the meter Any claim must be made within the warranty period The i SENS Inc will at its discretion repair or replace a defective meter or meter part that is cover...

Reviews: