34
‐
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the iBP-130, iBP-131,
iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9)
is used exceeds the applicable RF compliance level above, the
iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9)
should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9)
.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.