30
‐
Appendix I
Manufacturer’s declaration-electromagnetic immunity
The iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9) is intended for use in the
electromagnetic environment (for home healthcare) specified below.
The customer or the user of the iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9)
should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagn
etic
environment-
guidance
(for home
healthcare
environment)
Electrostatic
discharge(ESD) IEC
61000-4-2
Contact:
±
8 kV
Air
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV
Contact:
±
8 kV
Air
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV
Floors should
be wood,
concrete or
ceramic tile. If
floors are
covered with
synthetic
material, the
relative
humidity
should be at
least 30%
Electrical fast
transient/burst IEC
61000-4-4
+ 2kV for power supply lines
+ 1kV for input/output lines
Not applicable
Not applicable
Mains power
quality should
be that of a
typical home
healthcare
environment.
Surge IEC 61000-4-5
+ 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to
Not applicable
Not applicable
Mains power
quality should