Manufacturer’s declaration-electromagnetic immunity
The iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9) is intended for use in the electromagnetic
environment (for home healthcare) specified below.
The customer or the user of the iBP-130, iBP-131, iBP-132, iBP-133, iBP-143, iBP-1XX(XX=0-9)
should assure
that is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
(for home healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
:
0,15 MHz – 80 MHz
6 Vrms
:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
e)
10 V/m
80 MHz – 2,7 GHz
b)
80 % AM at 1 kHz
c)
Not applicable
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF
communications
equipment should be used no closer to
any part of the
iBP-130, iBP-131,
iBP-132, iBP-133, iBP-143,
iBP-1XX(XX=0-9)
including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1,2
√
P
d = 1,2
√
P
80MHz to 800 MHz
d = 2,3
√
P
800MHz to 2,7 GHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: