18
‐
3.4
Manual
Inflate
Function
If your systolic pressure is known to be more than 220 mmHg please follow procedures. Press
and hold the “Up" button to enter “Manual Inflate” mode before the cuff inflated to 180mmHg.
When cuff pressure reaches 290 mmHg, the cuff will be deflating and measurement starts.
If you do not wish to inflate to 290mmHg, you can release “Up" during the inflating process. The
air will be leaking and measurement starts.
When the cuff is inflated to 180mmHg,
press and hold the “Up" button to enter
“Manual Inflate” mode.