23
System Functions / Features
Dual Temperature Ranges**
Low Range may be set between 50
°
F and 99
°
F
This system incorporates two temperature range settings with independent set
temperatures. The High Range designated in the display by an “up” arrow or “High
Range”, and the Low Range designated in the display by a “down” arrow or “Low Range”.
High Range may be set between 80
°
F and 104
°
F
Freeze Protection is active in either range
These ranges can be used for various reasons, with a common use being a “ready to use”
setting vs. a “vacation” setting. The Ranges are chosen using the menu structure. Each
range maintains its own set temperature as programmed by the user. This way, when a
range is chosen, the spa will heat to the set temperature associated with that range.
Mode – Ready and Rest**
In order for the spa to heat, a pump needs to circulate water through the heater. The pump that
performs this function is know as the “heater pump.”
READY MODE
- Will circulate water every 1/2 hour, using Pump1 Low, in order to maintain a
constant water temperature, heat as needed, and refresh the temperature display. This is
known as “polling.” “Ready” will be displayed.
REST MODE
- Will only allow heating during programmed filter cycles. The temperature
display may not show a current temperature until the heater pump had been running for a
minute or two. “REST” will be displayed.
READY-IN-REST MODE -
Appears in the display if the spa is in Rest Mode and Jet 1 is
pressed. It is assumed that the spa is being used and will heat to set temperature. While Pump
1 High can be turned ON and OFF, Pump 1 Low will run until set temperature is reached, or 1
hour has passed. After 1 hour, the system will revert to Rest Mode. This mode can also be
reset by entering the Mode Menu and changing the Mode.
Time of Day**
Setting the time-of-day can be important in determining filtration times and other
background features
.
If power is interrupted to the system, Time-of-Day is not stored. The system will still operate
and all other user settings will be stored. If filtered cycles are required to run at a particular time
of day, resetting the clock will return the filter times to the actual programmed periods.
NOTE:
**See provided Spaside Control Operation
Guide for programming procedures.