Hydro-View User Guide HD0531 Rev 2.0.0 59
7.3
Edit Points Screen
Figure 45: Edit Calibration Points Screen
The main screen
Displays a list of the points currently used in the Calibration. Touch either the Unscaled or
Moisture box to edit the value. Touch the Include tick box to include / exclude the point from
the calculation of the coefficients. Excluded points are not shown on the graph. A line
highlighted in red indicates that data is missing, either because an Unscaled reading has
been taken from the sensor and is awaiting a laboratory moisture reading, or because it is the
blank line for a new point. Points with missing data are not included on the calculation of the
coefficients. A maximum of 20 points is allowed per calibration. A blank line is always present
at the bottom of the list for entering new points.
Scroll Up and Down buttons navigate the list of points. The list automatically scrolls if there
are more points than fit on the screen.
Averaging
Accesses the Remote Averaging Screen to obtain a representative material sample reading
from the sensor. Particularly important in a Batch Averaging application.
Graph
Shows the list of points on a graph to assist with choosing the best points for a good
Calibration.
Mode
Switches the Measurement Mode displayed in the calibration points list
OK
Accepts the edit of the Calibration points.
7.4
Averaging Screen
The Remote Averaging function is a convenient way to obtain an averaged reading of a
material as it passes the sensor (Figure 46). This is particularly important in batch averaging
applications. Remote Averaging works in two different ways, depending on the setting of the
sensor’s Digital Input.
Figure 46: Averaging Screen
Summary of Contents for HV04
Page 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...