4.
To change the features that are password protected, push
OK
.
5.
Enter the password of the current user.
Note: The security password is set the first time the security settings are changed.
6.
Select the features to be password protected.
7.
Push the
RIGHT
arrow to save the changes.
8.
To change the security password for the current user, push and then select Change Password.
Table 2 Password-protected features
Feature
Prevents
Calibration Intervals
Changes to the sensor calibration intervals
Calibration Type
Changes to the selected calibration types
Firmware Update
Changes to the firmware
Modify Log File
• Changes to a log after it is made and before it has started
• The active log from being stopped
Maintenance Intervals
Changes to the sensor maintenance intervals
Calibration
Calibration
Delete Log File from Device Removal of log files from the instrument
Sensor Settings
Changes to the sensor settings
Stability Check
Changes to the stability check criteria or stability check setting (enabled or
disabled)
Factory Reset
The user settings to be set back to the factory defaults.
Configure the sonde
Configure the sonde date, time and sounds. The date and time are saved with the measurement data
and the calibration history. The date and time are used to identify when sensor calibration or
maintenance is necessary. The sonde settings are saved on the sonde.
1.
Push until three or four dots show at the top of the screen.
2.
Push the
RIGHT
arrow until "Sonde HL" shows at the top of the display.
3.
Select Settings>Settings.
4.
Select an option, then push
OK
.
Option
Description
Set Time
Sets the date and time.
Sync with Surveyor
—copy the date and time from the Surveyor HL.
Manual
—enter the date and time.
Audio Enable
Sets sounds to on or off. The sonde gives audible beeps to identify conditions (e.g., when the
sonde comes on).
Configure the sensors
1.
Push until three or four dots show at the top of the screen.
2.
Push the
RIGHT
arrow until "Sonde HL" shows at the top of the display.
3.
Select Sensors.
4.
Select a sensor to configure.
The date of the last calibration, the calibration type, the calibration interval and the next
calibration date show.
12
English