CSI-B-1 Features
The condition sensor interface CSI-B-1 makes it possible to easily integrate the
following sensors into analog data acquisition systems:
- ContaminationSensor CS1000,
- AquaSensor AS1000,
- HYADCLab HLB1000.
These sensors determine two or four measured values which can be extracted via a
digital interface or an analog output.
In the case of analog output, the two or four measured values are either transferred
chronologically one after the other or the output is permanently set to a single
measurement channel.
In view of the fact that the data collection systems which are used often do not
enable any simple decoding of the chronologically coded analog signal, the CSI-B-1
is made available as an easy-to-use module which outputs the sensor data in the
form of four separate analog signals.
The connected sensor is recognized automatically – no configuration of the CSI-B-1
is required.
In addition, the module can be utilized as a miniature monitoring system. A readily
programmable switching output is available for each of the four measurement
channels which can be used to monitor whether a measurement parameter is within
or outside of a range or whether limit values are exceeded or not achieved.
The ConditionSensor Interface CSI-1-B can be parameterized through the USB
interface via hyperterminal.