CSI-B-1
HYDACLab - Reading out measured values
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 28 / 36
BeWa CSI-B-1 3424452b 2012-04-20 en-us.doc
2012-04-20
HYDACLab - Reading out measured values
If a HYDACLab is detected at the sensor input, then the CSI-B-1 outputs the
following analog signals:
Output
Value
OUT1
Temperature T (
−
25°C … +100°C)
Voltage U =
Voltage
U =
2 + 8 x (T + 25°C) / 125°C
Current I =
Current
I =
4 + 16 x (T + 25°C) / 125°C
Temperature
=
((voltage U
−
2) × 125 / 8)
−
25
Temperature
=
((current I
−
4) × 125 / 16)
−
25
OUT2
Rel. change in the dielectric constant rDK (
−
30 % … +30 %)
Voltage U =
Voltage
U =
2 + 8 x (rDK + 30%) / 60%
Current I =
Current
I =
4 + 16 x (rDK + 30%) / 60%
rDK =
rDK =
((voltage U
−
2) × 60 / 8)
−
30
rDK =
rDK =
((current I
−
4) × 60 / 16)
−
30
OUT3
Saturation level S (0 ... 100 %)
Voltage U =
Voltage
U =
2 + 8 x S / 100
Current I =
Current
I =
4 + 16 x S / 100
S
=
(voltage U
−
2) × 100 / 8
S
=
(current I
−
4) × 100 / 16