ContaminationSensor – CS 2000 series
HYDAC Filtertechnik GmbH
en
Page 33
BeWa CS2000 3117721e en.doc
2008-03-28
Press the “p” key and then the Return key to enter Parameter mode, in which you
can program the CS settings.
8.3.2 Parameter
mode
The following display appears after Parameter mode starts:
Parameter Mode:
x - Go back
t - Measuring Cycle Time
m - Measuring Mode
2 - Limits Mode M2
3 - Limits Mode M3
4 - Limits Mode M4
b - Set DIN-Bus Address and Name (for CS models with DIN-Bus module only)
e - Set IP Address (for CS models with Ethernet Interface only)
p - Pump Protection
o - PLC Output Format
r - RS232-Display Output Format
i - Current Output Format (for CS models with analog module only)
c - Check Current Output (for CS models with analog module only)
f - Flow Check
s - Sensor Check
a - Configure 4..20 mA Input Channels
h - Display AD Channels
d - Recover Factory Default Settings
z – Show Device Info
Press key (x, t, m, …) and Return!
8.3.3 Measurement interval
The measurement interval is set using
t =
Measuring time
(20 to 120 sec).
If a time shorter than 20 s is set (only for test purposes), this is not stored in the
EPROM, so that after a reset of the CS the last stored measurement interval set
within the permitted range (20 to 120 s) is restored.
It is not possible to set a time < 2 s or > 120 s.
Current measuring cycle time (s) set to: 20
New value for measuring cycle time (s) (press Return only to keep setting):
60
'Return
'
Current measuring cycle time (s) set to: 60