ContaminationSensor – CS 2000 series
HYDAC Filtertechnik GmbH
en
Page 28
BeWa CS2000 3117721e en.doc
2008-03-28
Ready relay
on
NC
NO
C
Ready relay
off
NC
NO
C
Relay 1
on
NO
C
Relay 1
off
NO
C
Relay 2
on
NO
C
Relay 2
off
NO
C
M4
Start or
result of the
check
measureme
nt after test
cycle time: a
value
≥
particular
upper limit
CS ready for
operation
CS not ready
for operation
Measurement
in progress
and in one or
more of the
last 5
measurement
s: a value >
particular
lower limit
In 5
successive
measurement
s:
all values
≤
particular
lower limit
or
measurement
stopped
After first
value is
present: Flow
error
Flow in
nominal
range
After end of
test cycle for
duration of a
check
measureme
nt
Test cycle
time has
elapsed
Off again if all
values <
particular
upper limit
Restart test
cycle time