ContaminationSensor – CS 2000 series
HYDAC Filtertechnik GmbH
en
Page 22
BeWa CS2000 3117721e en.doc
2008-03-28
5.2
Electrical connection of the CS
Before continuing, check whether the CS is suitable for your display/data capturing
system by carefully examining the technical specifications.
First connect the signal lines (pins/terminals 1 to 23).
Then connect the supply voltage.
CS starts measurement operation as soon as the supply voltage is applied.
5.3
Selecting a measurement point
1
The measurement point should be selected so that the measured volume
comes from a turbulent area with a good through-flow. For example: on a
pipe elbow, etc.
2
The CS has to be installed in the vicinity of the measurement point in order
to obtain precision-timed results.
3
While installing the measurement line, avoid creating a "siphon" trap for
sedimentation (particle deposits in the line).
A
B
2
A
B
1
3
A
B
2
1
WRONG WRONG
OK