background image

11

the next interface.

2 0 0 7 . 0 1 . 0 1

Tap

R

3 5 1 . 2

Store

Current tap:

0 1

Print

Re-test

Exit

1 4 : 0 2 : 5 1

Now, the dc resistance tester starts to display the resistance value of the

winding under test and such value gradually becomes stable. At this time,
press the Tap button to adjust and set the current tap value or winding name so
as to mark the measurement result. After the Tap button is pressed down, the

multi-functional key on the right becomes “▲”“▼”“

”“OK” and “Cancel”, and

the tap setting dialogue box is ejected. After setting, press “OK” button to save
it and press “Cancel” button to modify it. Then, press “Store” and “Print” button
to save or print the measurement result. In case of any doubt about the data,
press “Retest” button to carry out the measurement and calculation again.

When measuring the on-load voltage changing transformer, after the

testing data of one tapping position becomes stable, the on-load tapping
switch may be switched to the next tapping position, without discharging to
conduct the measurement over again. At this time, the resistance value of the
winding will change gradually until it becomes stable. You may also press
“Re-test” button to quickly refresh the data. Repeat the above-mentioned
procedure until all the tapping tests are completed.

Three-phase measurement:

The three-phase measurement applies to the windings of YN star

Summary of Contents for HTZZ-S10A

Page 1: ...manual You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column...

Page 2: ...and prevent the product or other relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out main...

Page 3: ...d the instruction manual to acquire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type an...

Page 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Page 5: ...I General 6 II Performance Index 7 III Panel Description 7 IV Testing and Operating Method 9 V Precautions 14 VI Troubleshooting 15 VII Transportation and Storage 15 VIII Unpacking and Inspection 16 I...

Page 6: ...hase measurement is recommended Caution The three phase measurement result doses not include the resistance value of the neutral lead Therefore it is suggested to at least verify one data using the si...

Page 7: ...Single phase 1A 5A 10A 20A Three phase 1A 5A 10A Measurement scope Single phase 1A 10 m 20 5A 1 m 4 10A 1 m 2 20A 1 m 1 Three phase 1A 10 m 6 each phase 5A 1 m 1 each phase 10A 1 m 0 6 each phase Mea...

Page 8: ...the first channel Uc and Uo are the input of the second channel 3 Panel printer Prints the output measurement results 4 LCD displays the measurement results and operation prompting information with t...

Page 9: ...easure 1 4 0 2 5 1 Set key press Set the current date and time View key press View the data stored including measurement results testing date tapping number and etc Mode key press Sets the measurement...

Page 10: ...nters into the interface as follows 2 0 0 7 0 1 0 1 Tap R 0 0 A Store The battery is charging please wait a moment Current tap 01 Print Re test Exit 1 4 0 2 5 1 At this time the current of the winding...

Page 11: ...tton to modify it Then press Store and Print button to save or print the measurement result In case of any doubt about the data press Retest button to carry out the measurement and calculation again W...

Page 12: ...the tester to enter into its measuring state and to start to supply power to the winding and then it enters into the interface as follows 2 0 0 7 0 1 0 1 Tap A 0 0 A B 0 0 A Store C 0 0 A The battery...

Page 13: ...rement result After the Tap button is pressed down the multi functional key on the right becomes OK and Cancel and the tap setting dialogue box is ejected After setting press the OK button to save it...

Page 14: ...g may be removed V Precautions 1 When measuring the no load voltage changing transformer be sure to exit from the discharge before switching the tapping switch Do not switch the tapping switch before...

Page 15: ...t VII Transportation and Storage Transportation When transporting the equipment it is suggested using the instrument packing wood boxes and damping materials of our company to avoid any unnecessary da...

Page 16: ...y shortage please make contact with our company We will give serve you as soon as possible IX Others The warranty period for the whole set of this product is one year and Three guarantees repair retur...