10
measuring current is selected, press “Measure” key-press to start the test.
2
Single-phase measurement:
The wiring method is shown in the figure below (Taking the measurement
as of R
AB
an example):
After the winding under test is wired and the suitable measuring current is
selected, press the “Measure” key-press to allow the tester to enter into its
measuring state and to start to supply power to the winding, and then it enters
into the interface as follows:
2 0 0 7 . 0 1 . 0 1
Tap
R
0. 0 A
Store
The battery is charging, please wait a moment
!
Current tap:
01
Re-test
Exit
1 4 : 0 2 : 5 1
At this time, the current of the winding starts to increase. If the charging
progress strip and the current display value stand still for a long time, it
indicates that the value of the resistance value to be measured is probably
beyond the measuring scope of the current, so it is not able to reach its set
value. At this time, press “Exit” key-press to return and re-select the current
and retry. When the set value is reached, it becomes constant. Then, enter into