98
5. Check OCT Scan
①
Select OCT Scan in MESURE Mode.
②
Select Macular 3D at Scan Pattern and change the OCT Sensitivity to Normal as below picture
and then turn the OCT Enface option to On.
③
Scan will start automatically when the alignment and focusing is correct, and let the Bscan, IR
Fundus and real time Enface screen to be displayed as below picture using Optimize function,
-
Check if the Split Focus is aligned as straight line in IR FUNDUS screen.
-
Check if the Retina image of the Bscan screen is in between the green guidelines.
-
Check if the real time enface video marked with red rectangle shows grid pattern.