97
HOCT-1F/1
4. Check Fundus Camera
①
Select FUNDUS Mode and measure Fundus.
②
Check if there are asymmetric areas as right picture from the measured images.
-
If there are some, re-boot the device and measure again.
③
Check if there are any blurriness or smear as below picture from the scanned images.
-
If blurriness or smear appears, clean the finger prints or stain on the objective lens.
NG
OK