
45
HOCT-1F/1
6.4.
Anterior segment image operation (optional)
6.4.1.
Preparation for anterior segment operation
1.
Check the lens surface of anterior segment adapter is clean.
2.
Thread anterior segment adapter (A) to objective lens holder (B), and check there is no tilting or
misalignment of adopter. Don’t lock it too hard.
3.
Check anterior headrest rubber is clean.
4.
Hang anterior headrest rubber to headrest of instrument (C) first, then press anterior headrest rubber
to be fixed firmly.
5.
Check the mounted anterior headrest rubber does not incline.
Summary of Contents for HOCT-1F/1
Page 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1F 1 USER MANUAL ...
Page 4: ...4 9 EMC INFORMATION 103 10 SERVICE INFORMATION 105 11 SOFTWARE LICENSE AGREEMENTS 106 ...
Page 18: ...18 Rear View ...
Page 62: ...62 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Page 73: ...73 HOCT 1F 1 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Page 102: ...102 8 3 Drawings of System ...