26
Car
e and Cleaning
6.2.2 Receiver
Unit
General Comments
•
Always keep the external surfaces clean and free of dirt and fl uids using a
clean dry cloth to remove.
•
Clean the charging contacts regularly with a dry cloth.
•
Ensure that the charging contacts are completely dry following the cleaning
and disinfecting procedure.
Cleaning & Disinfecting Procedure
Following patient examination, clean and disinfect the receiver’s exterior surface
as described below:
1. Wipe any fl uids from the surface of the unit using a clean dry cloth.
2. Wipe the receiver with a cloth dampened in 70% Isopropyl Alcohol avoiding
connector sockets and charging pins.
3. Completely dry the receiver with a clean, dry lint free cloth.
4. If the receiver has been contaminated with blood disinfect the contaminated
area using a cloth dampened in Sodium Hypochlorite solution @ 10000ppm.
5. After two minutes wipe this area with a cloth dampened in water to remove
residue and then dry with a lint free cloth.