No battery indicator LEDs illuminated
POSSIBLE CAUSE SOLUTION
RF Interference
Occasionally, the battery indicator LEDs may show
absence of charge. As long as the transducers have been
properly docked, with the charge bay LEDs showing that
charging is in progress for a minimum of 15 minutes, the
transducers should operate normally. (Refer to Section 3.2)
8.1 Manually Switching off the Transducers
If the green LED of the ultrasound or Toco transducer remain illuminated or the
transducers lock up, the following procedure must be adhered to:
Magnet
Switch
Position of Magnet and Switch
Inside the housing by the upper right
edge of the receiver unit there is a
magnet. This is used to manually switch
off the transducer. The transducers
have an internal magnetic switch, which
is mounted within the transducer and
positioned under the C of “TOCO” and
the S of "US" respectively