WARNING: These requirements must be met when Sonicaid
™
Freedom is connected to any other electrical equipment.
1
Medical equipment must comply with IEC60601-1/EN60601-1, or
equivalent.
2
The confi gured system must comply with the requirements of
IEC60601-1:2005; clause 16.
4.1 Receiver
Connection
The receiver unit should be plugged into the same mains supply
circuit as other equipment in use on the same patient. Do not
make any modifi cations to the power supply of the receiver and
transducer units.
•
Connect the interface cable which is clearly identifi ed, from output of the
receiver unit to the interface input of the fetal monitor.
•
Connect the receiving antenna and select RF channel (See Section 8.3)
•
Dock the transducers with the receiver powered off.
•
Connect the mains supply lead from the receiver unit to mains supply. This
will pair the receiver and transducers.
•
Charge the transducer batteries
•
Switch on the Fetal Monitor (refer to its Instructions for Use)
The system is ready for monitoring.
4.2 Testing the Basic Functions
A functional test should be performed after initial installation, following each step
below:
1. Remove TOCO-transducer, while the US-transducer is charging:
•
charge bay LED and battery indicator LEDs turn off
•
LED in TOCO-transducer starts fl ashing
•
no further action (no signal transmission, because the US-transducer is not
running)