25
Car
e and Cleaning
Method 2 (Medium Risk)
Remove the bumpers before cleaning.
1. Remove soiling and clean the transducers and the internal and external
surfaces of bumpers with a mild detergent and then rinse in water.
2. Completely dry the transducers and the inside and external surfaces of the
bumper.
3. Wipe transducers and bumpers with a cloth dampened in Sodium
Hypochlorite (1000ppm).
4. After two minutes rinse with water and then dry with a clean lint free cloth.
5. Re-attach the bumper ensuring that they are assembled correctly (see
pictures
below).
Method 3 (High Risk)
Warning: Sodium Hypochlorite @ 10000 ppm for disinfecting
should only be used in situations described in the High Risk
defi nition. Unnecessary use of this concentrated solution for
Low and Medium risk situations may result in damage to the
transducer over time.
Remove the bumpers before cleaning.
1. Remove soiling and clean the transducers and the internal and external
surfaces of the bumpers with a mild detergent and then rinse in water.
2. Completely dry the transducers and the inside and external surfaces of the
bumpers.
3. Wipe transducers and bumpers with a cloth dampened in Sodium
Hypochlorite (10000ppm).
4. After two minutes rinse with water and then dry with a clean lint free cloth.
5. Re-attach the bumper ensuring that they are assembled correctly (see
pictures below).
Warning: Before docking the transducers onto the Receiver,
ensure that the transducers are dry and that the receiver has
been cleaned following the procedure below. If this is not
followed, damage to transducer will result.