Press
(INCREASE) or
(DECREASE) repeatably to scroll through
the available channel names for an appropriate choice. Choices for channel
names are: CH1, CH2, CH3, CH4, LOAD1, LOAD2, LOAD3, LOAD4, DISP1,
DISP2, DISP3, DISP4, FLOW, STAB, VERT, HORZ, PRES1, PRES2, PRES3, PRES4,
VOLU1, VOLU2, VOLU3 and VOLU4.
NOTE:
It is important to assign an appropriate name when naming
channels because the HM-3000 will choose a unit of measure based on the
name assigned. For example, if you are configuring a load cell, name the
channel Load 1, 2, 3 or 4. This will ensure that the channel reads in lbs (kN).
The same applies to Displacement (DISP 1, 2, 3 or 4) reading in inch (mm)
and Pressure (PRES1, 2, 3 or 4) reading in psi (kPa) etc.
Once a name has been chosen for a channel Press
to return to the
previous screen. You should see this screen (LOAD1 was chosen as name):
Sensor Min
The Sensor Zero (SENSOR MIN) is not configurable and is set at 0.
Sensor Max
Press
(SENSOR MAX) to set the Sensor Maximum value.
You should see this screen:
“CH1” CONFIGURATION
* NAME =
“CH1”
* INCREASE
* DECREASE
CH1 CONFIGURATION
* NAME =
“LOAD1”
* SENSOR MIN =
0
* SENSOR MAX =
“VALUE”
18
“CH1” MAX SETUP
* SENSOR MAX =
“000000”
* INCREASE
* DECREASE