5. Troubleshooting
Message/Symptom
Cause
Treatment
Can't Find Device File
Press [ENTER] to do new device.
There is no device or RFID Tag
Insert the cartridge or check whether
RFID Tag is attached or not.
There are two or more devices or
RFID Tags
Remove RFID Tag except that need to
test.
Unknown Device. Press [ENTER].
Insert Appropriate Device.
There are no test(ddb) files.
Register the test(ddb) files using USB.
Please refer to the “3.5.2 New Device
Registration” section.
Unknown Specimen. Press
[ENTER]. Retest with New Device.
Specimen Type Recognition Error.
Try it with another device and exact
specimen. If the same error is occurred,
request A/S.
Date Expired!
Expiry Date of devices is passed.
Test again with the cartridge that has
not expired.
Can't Find Window 1/2 Control
Line. Press [ENTER] to Previous.
Internal accuracy control error.
The specimen does not move
with membrane figure.
Try it with another device to retest.
Log in Error.
Log in with the unregistered
card.
Use after registering the ID card using
the supervisor card.
Abnormality of System Icon .
Error with Camera Module
Connection, Control Board
Connection and Transmission of
specimen measurement
command, RFID Module
initialization and connection,
Thermal Printer Connection, or
miscellaneous system.
Turn off the device and leave it off for
15 seconds, and reboot. If the same
error is occurred, request A/S
Abnormality
of IQC Test Icon
IQC Test Result is Fail.
Restart IQC Test. If an error occurs,
proceed with the calibration. After
calibration, test IQC again.
IQC Test duration has past.
Reset IQC Test duration from
Parameters Menu.
Can't find the corresponding
marker(IQC Test).
The IQC device line is damaged.
If the same error is occurred, request
A/S and receive the new IQC device.
Fail calibration and Press
[ENTER].
In case of auto calibration failure.
If the same error is occurred, request
A/S.
Can't find the corresponding
marker(QC Test).
When testing with the device
and other coding device.
Check the device and corresponding
coding device you want to test.