1. Select “QC test setting” menu, then press button.
2. Use
on setting menu and
on selected menu for adjusting test condition.
3. Once press
button after finishing all test condition setting, the setting will be saved. If saving is not
wanted, press
button to move back to previous menu.
<QC TEST SETTINGS 메뉴화면> <QC TEST SETTINGS 변경메뉴>
3.4 QC Test Settings
Menu
Default
Option
Description
Set QC Test
QC Frequency
None
8hr, Daily, weekly,
Monthly
Setting menu for checking period of QC test.
Sample
2
2,1
Determining menu for number of QC sample.
SD
X3 SD
X3 SD, X2 SD
Setting menu for valid range of QC test result.
Set IQC TEST
Frequency
None
8hr, Daily, weekly,
Monthly
Setting menu for checking period of IQC test.
IQC Fail
Use System
Use system,
Block System
Activate or deactivate equipment setting menu what if IQC
test result is fail.
<QC TEST SETTINGS Screen> <QC TEST SETTINGS Manu Screen>