www.humancaregroup.com
15
ENGLISH
Description of parts
Hand control
The hand control allows you to control the
various functions of the lift.
Depending on the version, two types of hand
control are available.
1. Battery charge
The battery charge indicator has three status
indications. The battery charge indicator is
yellow or green until switched off (2 minutes
after use).
For a full description, see page 18.
2. Service indicator
The service indicator (yellow indicator)
flashes when it is time to carry out a service.
The default setting is every 12 months /
8000 lifting cycles (as recommended by the
EN10535 standard), whichever comes first.
When it is time for a service, the service
indicator stays on for 2 minutes after use. It
is then switched off to save battery life. The
system is still functional when the service
indicator is on.
3. Overload indicator
When there is an overload, according to the
predefined power interruption, the two LED
lights will flash for 10 seconds.
4. Up
This button raises the lift arm, to help a
client into a standing-up position.
5. Down
This button lowers the lift arm, to help a
client into a seated position.
1
2
3
4
8
9
5
6
7
Diana II Comfort MSB
Diana II Comfort ESB
6. Leg adjustment, open
The legs can be opened so that the lift can
be brought closer to a client sitting in a chair
or lying on the floor. This text is also valid for
the Diana II.
7. Leg adjustment, close
The legs can be closed to make it easier to
maneuver and negotiate narrow passage
-
ways (e.g. doorways).
8. Tilting position, forwards
9. Tilting position, backwards
Summary of Contents for 90559